FDA Adverse Event Malfunction Summary report: N

CTF03, 5X100 KII FIOS Z-THR 6/BX

MDR report key: 14342501 · Received May 9, 2022

Report

Report Number
2027111-2022-00597
Event Type
Malfunction
Date Received
May 9, 2022
Date of Event
December 16, 2021
Report Date
June 7, 2022
Manufacturer
APPLIED MEDICAL RESOURCES
Product Code
GCJ
UDI-DI
00607915123550
PMA / PMN Number
K041795
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

NO PRODUCT IS BEING RETURNED TO APPLIED MEDICAL FOR EVALUATION. A FOLLOW-UP REPORT WILL BE PROVIDED UPON COMPLETION OF THE INVESTIGATION.

Additional Manufacturer Narrative · 0

THE EVENT UNIT WAS NOT RETURNED TO APPLIED MEDICAL FOR EVALUATION. AS THE EVENT UNIT WAS NOT RETURNED, APPLIED MEDICAL IS UNABLE TO DETERMINE IF THE EVENT UNIT EXHIBITED ANY NON-CONFORMANCES THAT COULD HAVE CONTRIBUTED TO THE REPORTED EVENT. IN THE ABSENCE OF THE EVENT UNIT, IT IS DIFFICULT TO DETERMINE IF THE REPORTED EVENT WAS CAUSED BY A MANUFACTURING NON-CONFORMANCE OR CIRCUMSTANTIAL FACTORS AT THE TIME OF USE. APPLIED MEDICAL HAS REVIEWED THE DETAILS SURROUNDING THE EVENT AND IS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE EVENT. THE PROBABILITY AND CRITICALITY OF HARM RESULTING FROM THIS FAILURE HAVE BEEN EVALUATED AND WERE FOUND TO BE AT AN ACCEPTABLE LEVEL.

Description of Event or Problem · 0

EVENT 1 OF 5: (B)(4) (REPORT # 2027111-2022-00557). EVENT 2 OF 5: (B)(4) (REPORT # 2027111-2022-00558). EVENT 3 OF 5: (B)(4). EVENT 4 OF 5: (B)(4). EVENT 5 OF 5: (B)(4). PROCEDURE PERFORMED: UNKNOWN. EVENT DESCRIPTION: A FAULT HAS OCCURRED WITH ITEM SKU (B)(4), REF: CTF03, LOT: 1411612, SUPPLIED FROM THE NDC. ON (B)(6) 2021 DURING THE THORACIC PART OF THE PROCEDURE IT WAS NOTICED THAT THE TIP OF 5MM TROCAR WAS MISSING. PRESUMED IN PATIENT BUT NOT FOUND DUE TO SMALL SIZE AND PLASTIC , NOT SHOW ON XRAY. EXPIRY DATE OF PRODUCT MARCH 2024. NO OF INSTANCES: 2-5. ADDITIONAL INFORMATION RECEIVED BY EMAIL FROM APPLIED MEDICAL REPRESENTATIVE ON 21APR22: I HAVE TRIED TO FOLLOW UP WITH THIS [ADDITIONAL INFORMATION REGARDING THE OTHER EVENTS] AND NO ONE WILL RESPOND TO ME.

Description of Event or Problem · 0

PROCEDURE PERFORMED: UNKNOWN EVENT DESCRIPTION: COMPLAINT 1 OF 5: (B)(4) [MFR REPORT # 2027111 - 2022 - 00557] COMPLAINT 2 OF 5: (B)(4) [MFR REPORT # 2027111 - 2022 - 00558] COMPLAINT 3 OF 5: (B)(4) [MFR REPORT # 2027111 - 2022 - 00596] COMPLAINT 4 OF 5: (B)(4) [MFR REPORT # 2027111 - 2022 - 00597] COMPLAINT 5 OF 5: (B)(4) [MFR REPORT # 2027111 - 2022 - 00598] A FAULT HAS OCCURRED WITH ITEM SKU 153343, REF: CTF03, LOT: 1411612, SUPPLIED FROM THE NDC. ON (B)(6) 2021 DURING THE THORACIC PART OF THE PROCEDURE IT WAS NOTICED THAT THE TIP OF 5MM TROCAR WAS MISSING. PRESUMED IN PATIENT BUT NOT FOUND DUE TO SMALL SIZE AND PLASTIC , NOT SHOW ON XRAY. EXPIRY DATE OF PRODUCT MARCH 2024. NO OF INSTANCES: 2-5. ADDITIONAL INFORMATION RECEIVED BY EMAIL FROM APPLIED MEDICAL REPRESENTATIVE ON 21APR22: I HAVE TRIED TO FOLLOW UP WITH THIS AND NO ONE WILL RESPOND TO ME. PATIENT STATUS: NO PATIENT INJURY HAS BEEN REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1839743 CTF03, 5X100 KII FIOS Z-THR 6/BX LAPAROSCOPE, GENERAL & PLASTIC SURGERY GCJ APPLIED MEDICAL RESOURCES CTF03 1411612 00607915123550

Patients

Seq Age Sex Outcome Treatment
1 Prefer Not To Disclose