DREAMSTATION BIPAP AVAPS30
Report
- Report Number
- 2518422-2022-19104
- Event Type
- Injury
- Date Received
- May 9, 2022
- Date of Event
- April 21, 2022
- Report Date
- June 23, 2023
- Manufacturer
- RESPIRONICS, INC.
- Product Code
- BZD
- UDI-DI
- 00606959045651
- PMA / PMN Number
- K102465
- Removal / Correction Number
- RES 88058
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- 003
Narratives
THE MANUFACTURER PREVIOUSLY SUBMITTED MDR 2518422-2022-19104-1 WITH INCORRECT SECTIONS B1, B2, H1, H6. CORRECTIONS TO PREVIOUS MDR ARE MADE IN THIS REPORT AS FOLLOWS. SECTION B1 WAS CORRECTED TO ADVERSE EVENT AND PRODUCT PROBLEM. (ONLY PRODUCT PROBLEM WAS CHECKED IN PREVIOUS MDR). SECTION B2 WAS CORRECTED TO OTHER SERIOUS OR IMPORTANT MEDICAL EVENTS AND REQUIRED INTERVENTION TO PREVENT PERMANENT IMPAIRMENT/DAMAGE. (PREVIOUSLY IT WAS BLANK). SECTION H1 WAS CHANGED FROM MALFUNCTION TO SERIOUS INJURY. SECTION H6- HEALTH IMPACT CODES WERE UPDATED.
THE MANUFACTURER PREVIOUSLY REPORTED AN ALLEGATION OF AN ISSUE RELATED TO SOUND ABATEMENT FOAM. ADDITIONAL INFORMATION WAS RECEIVED AND SECTION B5 SHOULD BE REPORTED AS: THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A BIPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT ALLEGED TO HAVE BLADDER CANCER. THERE WAS NO MEDICAL INTERVENTION REQUIRED BY THE PATIENT. THE REPORTED EVENT OF BLADDER CANCER AND ITS REPORTED SEVERITY WAS REVIEWED BY THE MANUFACTURE'S CLINICAL EXPERT. THIS EVENT IS ASSESSED AS NOT RELATED TO THE DEVICE IN THIS CASE. BASED ON THE AVAILABLE INFORMATION, THE MANUFACTURE CONCLUDES NO FURTHER ACTION IS NECESSARY. THERE WAS NO MEDICAL INTERVENTION REQUIRED BY THE PATIENT. THE DEVICE HAS NOT YET RETURNED TO THE MANUFACTURER FOR EVALUATION. THE PATIENT HAS SWAPPED OUT THE DEICE AT THE DISTRIBUTOR AND AT THIS TIME, NO FURTHER INVESTIGATION CAN BE PERFORMED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP REPORT WILL BE FILED. SECTION(S) B1, B2, HAS CHANGED RELATED TO THE COMPLAINT CHANGING FROM THE REPORTED ADVERSE EVENT TO A PRODUCT PROBLEM. SECTION H1 HAS CHANGED TO REFLECT A MALFUNCTION. SECTION H6 HEALTH EFFECT- IMPACT CODE, TYPE OF INVESTIGATION FINDINGS AND INVESTIGATION CONCLUSIONS HAS BEEN UPDATED.
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A BIPAP DEVICE'S SOUND ABATEMENT FOAM BECAME DEGRADED AND CAUSED BLADDER CANCER. THE PATIENT HAS ALLEGED TO SEEING A WHITE POWDER IN THE TANK AND HUMIDIFIER. THE PATIENT DID REPORT TO RECEIVE MEDICAL INTERVENTION AND IS TAKING A PRESCRIBED MEDICATION, BCG. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1804774 | DREAMSTATION BIPAP AVAPS30 | VENTILATOR, NON-CONTINUOUS (RESPIRATOR) | BZD | RESPIRONICS, INC. | DSX1130H11C | 00606959045651 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other| R |