DREAMSTATION CPAP PRO
Report
- Report Number
- 2518422-2022-19032
- Event Type
- Malfunction
- Date Received
- May 9, 2022
- Date of Event
- April 15, 2022
- Report Date
- July 20, 2023
- Manufacturer
- RESPIRONICS, INC.
- Product Code
- BZD
- PMA / PMN Number
- K131982
- Removal / Correction Number
- RES 88058
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER PREVIOUSLY RECEIVED INFORMATION ALLEGING VISUALIZATION OF PARTICLES RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. REPEATED ATTEMPTS TO HAVE THE DEVICE AND COMPONENTS RETURNED FOR EVALUATION AND INVESTIGATION WERE UNSUCCESSFUL. THE MANUFACTURER BELIEVES THEY WILL BE UNABLE TO GATHER ADDITIONAL INFORMATION. THE MANUFACTURER IS SUBMITTING A FINAL REPORT AT THIS TIME. IF PERTINENT INFORMATION BECOMES AVAILABLE TO THE MANUFACTURER AT A LATER DATE, AN ADDENDUM TO THIS FINAL REPORT WILL BE FILED.
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE/RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. IN PREVIOUS REPORT MDR 2518422-2022-19032-1, THIS SECTION WAS WRONGLY CAPTURED, CORRECTION TO THIS SECTION SHOULD BE: THE MANUFACTURER PREVIOUSLY REPORTED AN ALLEGATION OF AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE MANUFACTURER RECEIVED INFORMATION ALLEGED VISUALIZATION OF PARTICLES. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. THE DEVICE WAS RETURNED TO THE MANUFACTURER'S SERVICE CENTER FOR FURTHER EVALUATION. THE DEVICE WAS EVALUATED. THERE WAS MENTION OF FOAM DEGREDATION SEEN DURING EVALUATION. THE INTERNAL ASPECT OF THE DEVICE WAS INSPECTED. THE DEVICE POWERED ON AND AIRFLOW WAS CONFIRMED. THE DEVICE'S DOWNLOADED LOGS WERE REVIEWED BY THE MANUFACTURER. THERE WERE NO ERRORS FOUND. THE MANUFACTURER CONCLUDES THAT THEY COULD CONFIRM THE CUSTOMER'S ALLEGATION AND THERE WAS VISIBLE FOAM DEGRADATION AND UNIT SCRAPPED DUE TO 400 MODEL. IN PREVIOUS REPORT MDR 2518422-2022-19032-1, SECTION D8, D9 WAS NOT CAPTURED, G3 DATE WAS WRONGLY CAPTURED SHOULD BE 10/10/2022, H2 SECTION WAS NOT COMPLETELY CAPTURED, H3 WAS NOT CAPTURED, SECTION H6 WAS INCORRECTLY CAPTURED, CORRECTED ALL THESE SECTIONS AND UPDATED IN THIS REPORT.
THE MANUFACTURER RECEIVED INFORMATION ALLEGING VISUALIZATION OF PARTICLES RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1460157 | DREAMSTATION CPAP PRO | VENTILATOR, NON-CONTINUOUS (RESPIRATOR) | BZD | RESPIRONICS, INC. | DSX400H11W |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |