FDA Adverse Event Injury Summary report: N

FAST-FIX 360 CURVED NDL DELIVERY SYS

MDR report key: 14337399 · Received May 9, 2022

Report

Report Number
1219602-2022-00664
Event Type
Injury
Date Received
May 9, 2022
Date of Event
April 19, 2022
Report Date
August 8, 2022
Manufacturer
SMITH & NEPHEW, INC.
Product Code
GAT
UDI-DI
00885554023077
PMA / PMN Number
K092508
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INTERNAL COMPLAINT REFERENCE: (B)(4).

Additional Manufacturer Narrative · 0

H3, H6: THE REPORTED DEVICE WAS NOT RETURNED TO THE DESIGNATED COMPLAINT UNIT FOR INDEPENDENT EVALUATION, THUS VISUAL INSPECTION AND FUNCTIONAL TESTING COULD NOT BE PERFORMED. A REVIEW OF DEVICE RECORDS SHOWED THERE WERE NO INDICATIONS TO SUGGEST THAT THE PRODUCT DID NOT MEET MANUFACTURING SPECIFICATION UPON RELEASE FOR DISTRIBUTION. A COMPLAINT HISTORY REVIEW FOUND SIMILAR REPORTED EVENTS. THE INSTRUCTIONS FOR USE WAS REVIEWED AND FOUND TO INCLUDE CONDITIONS OF OFF LABEL USE AND TECHNIQUE SPECIFICS, AS WELL AS PRECAUTIONS AND WARNINGS RELATED TO THE USE OF THE DEVICE. A RISK MANAGEMENT REVIEW FOUND THAT THE REPORTED FAILURE WAS DOCUMENTED APPROPRIATELY, AND THERE WERE NO INDICATIONS TO SUGGEST THE ANTICIPATED RISK IS NOT ADEQUATE. THERE WAS NO RELATIONSHIP FOUND BETWEEN THE DEVICE AND THE REPORTED EVENT. THE COMPLAINT WAS NOT CONFIRMED. PLEASE REFER TO THE INSTRUCTIONS FOR USE FOR RECOMMENDATIONS ON PROPER USE OF THE DEVICE AND POTENTIAL TROUBLESHOOTING METHODS TO PREVENT FUTURE REOCCURRENCE OF THE REPORTED EVENT. NO CONTAINMENT OR CORRECTIVE ACTIONS ARE RECOMMENDED AT THIS TIME. IF THE PRODUCT ASSOCIATED WITH THIS EVENT IS RETURNED AT A FUTURE DATE, THIS INVESTIGATION WILL BE REOPENED FOR EVALUATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING MENISCUS REPAIR, THE FAST FIX WITH LOT NUMBER 2082656 FAILED, T2 DID NOT DEPLOY. THE PROCEDURE WAS COMPLETED WITH A NON-SIGNIFICANT SURGICAL DELAY USING A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1408598 FAST-FIX 360 CURVED NDL DELIVERY SYS SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE GAT SMITH & NEPHEW, INC. 72202468 2082656 00885554023077

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other