FAST-FIX 360 CURVED NDL DELIVERY SYS
Report
- Report Number
- 1219602-2022-00664
- Event Type
- Injury
- Date Received
- May 9, 2022
- Date of Event
- April 19, 2022
- Report Date
- August 8, 2022
- Manufacturer
- SMITH & NEPHEW, INC.
- Product Code
- GAT
- UDI-DI
- 00885554023077
- PMA / PMN Number
- K092508
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
INTERNAL COMPLAINT REFERENCE: (B)(4).
H3, H6: THE REPORTED DEVICE WAS NOT RETURNED TO THE DESIGNATED COMPLAINT UNIT FOR INDEPENDENT EVALUATION, THUS VISUAL INSPECTION AND FUNCTIONAL TESTING COULD NOT BE PERFORMED. A REVIEW OF DEVICE RECORDS SHOWED THERE WERE NO INDICATIONS TO SUGGEST THAT THE PRODUCT DID NOT MEET MANUFACTURING SPECIFICATION UPON RELEASE FOR DISTRIBUTION. A COMPLAINT HISTORY REVIEW FOUND SIMILAR REPORTED EVENTS. THE INSTRUCTIONS FOR USE WAS REVIEWED AND FOUND TO INCLUDE CONDITIONS OF OFF LABEL USE AND TECHNIQUE SPECIFICS, AS WELL AS PRECAUTIONS AND WARNINGS RELATED TO THE USE OF THE DEVICE. A RISK MANAGEMENT REVIEW FOUND THAT THE REPORTED FAILURE WAS DOCUMENTED APPROPRIATELY, AND THERE WERE NO INDICATIONS TO SUGGEST THE ANTICIPATED RISK IS NOT ADEQUATE. THERE WAS NO RELATIONSHIP FOUND BETWEEN THE DEVICE AND THE REPORTED EVENT. THE COMPLAINT WAS NOT CONFIRMED. PLEASE REFER TO THE INSTRUCTIONS FOR USE FOR RECOMMENDATIONS ON PROPER USE OF THE DEVICE AND POTENTIAL TROUBLESHOOTING METHODS TO PREVENT FUTURE REOCCURRENCE OF THE REPORTED EVENT. NO CONTAINMENT OR CORRECTIVE ACTIONS ARE RECOMMENDED AT THIS TIME. IF THE PRODUCT ASSOCIATED WITH THIS EVENT IS RETURNED AT A FUTURE DATE, THIS INVESTIGATION WILL BE REOPENED FOR EVALUATION.
IT WAS REPORTED THAT DURING MENISCUS REPAIR, THE FAST FIX WITH LOT NUMBER 2082656 FAILED, T2 DID NOT DEPLOY. THE PROCEDURE WAS COMPLETED WITH A NON-SIGNIFICANT SURGICAL DELAY USING A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1408598 | FAST-FIX 360 CURVED NDL DELIVERY SYS | SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE | GAT | SMITH & NEPHEW, INC. | 72202468 | 2082656 | 00885554023077 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |