COBAS INTEGRA 800
Report
- Report Number
- 1823260-2009-03481
- Event Type
- Malfunction
- Date Received
- May 15, 2009
- Date of Event
- April 24, 2009
- Report Date
- May 15, 2009
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JJE
- PMA / PMN Number
- K951595
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE FIELD SERVICE REPRESENTATIVE COULD NOT DETERMINE A CAUSE AND NOTED, HE CHECKED THE SAMPLE AND FOUND IT TO BE CLEAR. HE PERFORMED A CHECK TEST WITH HIGH RESULTS AND REPLACED THE ST2 PROBE TO CORRECT THIS. QC AND PATIENT SAMPLES WERE PERFORMED TO VERIFY THE ANALYZER PERFORMANCE.
THE USER RECEIVED AN ERRONEOUS ALP RESULT FOR ONE PATIENT SAMPLE. THE INITIAL RESULT WAS 139 U PER L, REPEAT RESULTS WERE 118 AN 275 U PER L. THE USER THEN RAN THE SAMPLE IN A MICROCUP FIVE TIMES TO CHECK THE ALP PRECISION WITH THE FOLLOWING RESULTS: 119, 120, 119, 119 AND 116 U PER L. ALL TESTING WAS PERFORMED ON THE SAME DAY FROM THE SAME SAMPLE TUBE. THE USER STATED THEY THEN SENT THE SAMPLE OUT AND IT RECOVERED 119 U PER L SEVERAL TIMES, AND THIS RESULT WAS REPORTED. PATIENT NOT ADVERSELY AFFECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COBAS INTEGRA 800 | CLINICAL CHEMISTRY ANALYZER - JJE | JJE | ROCHE DIAGNOSTICS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |