FDA Adverse Event Injury Summary report: N

BLT Ø3.3MM NC, SLACTIVE® 10MM, TIZR, NTP

MDR report key: 14318572 · Received May 7, 2022

Report

Report Number
0009613348-2022-40650
Event Type
Injury
Date Received
May 7, 2022
Date of Event
February 10, 2020
Report Date
May 6, 2022
Manufacturer
INSTITUT STRAUMANN AG
Product Code
DZE
UDI-DI
07630031706621
PMA / PMN Number
K140878
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE BATCH NUMBER COULD NOT BE VERIFIED DUE TO INCOMPLETE OR MISSING INFORMATION AND / OR PRODUCT FROM THE CUSTOMER. OUR MANUFACTURING Q-SYSTEM ASSURES THAT PRODUCTION AND PROCESS CONTROLS ARE IN PLACE TO ENSURE THAT BATCHES CONFIRM TO THE APPLICABLE SPECIFICATIONS BEFORE THEY ARE DISTRIBUTED. THE LOSS OF AN ENDOSSEOUS DENTAL IMPLANT AFTER SUCCESSFUL OSSEOINTEGRATION AND RESTORATION IS A KNOWN INHERENT RISK OF THE PROCEDURE. IMPLANTS MAY HAVE TO BE REMOVED IN CASE ONE OR MORE OF THE IMPLANT SUCCESS CRITERIA ARE NOT MET. IMPLANT SUCCESS CRITERIA ACCORDING TO BUSER ET AL. (1991) ARE: 1. ABSENCE OF PERSISTENT SUBJECTIVE COMPLAINTS SUCH AS PAIN, FOREIGN BODY SENSATION AND /OR DYSESTHESIA. 2. ABSENCE OF A RECURRENT PERI-IMPLANT INFECTION WITH SUPPURATION. 3. ABSENCE OF IMPLANT MOBILITY. 4. ABSENCE OF A CONTINUOUS RADIOLUCENCY AROUND THE IMPLANT. THE MANUFACTURER'S TREND ANALYSIS CONFIRMS THAT THE REPORTED FAILURE RATE ASSOCIATED WITH ITS DENTAL IMPLANTS IS BELOW THE EXPECTED FAILURE RATE FOR THIS TREATMENT AS PUBLISHED IN THE SCIENTIFIC LITERATURE.

Description of Event or Problem · 0

THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)96) 2015 IN FDI 26. ON (B)(6) 2020, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1251180 BLT Ø3.3MM NC, SLACTIVE® 10MM, TIZR, NTP ENDOSSEOUS DENTAL IMPLANT DZE INSTITUT STRAUMANN AG SLActive Roxolid BL 07630031706621

Patients

Seq Age Sex Outcome Treatment
1 62 YR Female Required Intervention