FDA Adverse Event Malfunction Summary report: N

HEATER-COOLER SYSTEM 3T

MDR report key: 14317303 · Received May 7, 2022

Report

Report Number
9611109-2022-00229
Event Type
Malfunction
Date Received
May 7, 2022
Date of Event
April 8, 2022
Report Date
May 6, 2022
Manufacturer
LIVANOVA DEUTSCHLAND
Product Code
DWC
PMA / PMN Number
K191402
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THERE WAS NO KNOWN PATIENT INVOLVEMENT LIVANOVA (B)(4) IMPLEMENTED A FIELD SAFETY NOTICE FOR DISINFECTION AND CLEANING OF HEATER-COOLER DEVICES. THE Z NUMBER IS Z-2076/2081-2015. LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T. THE INCIDENT OCCURRED IN (B)(6). IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 0

H.10: ADDITIONAL DOCUMENTATION REGARDING MICROBIAL TESTS CONDUCTED ON THE MACHINE WERE SHARED WITH LIVANOVA. ANALYSIS REVEALED THAT THE FIRST POSITIVE WATER SAMPLE WAS FOUND IN SEPTEMBER 2021 AND UNTIL JUNE 2022 THE DEVICE NEVER TESTED NTM NEGATIVE. THE MOST LIKELY ROOT CAUSE OF THE REPORTED CONTAMINATION CAN BE RELATED TO THE LOCATION WHERE DEVICE IS USED/STORED. HOWEVER, NO ADDITIONAL INFORMATION IS AVAILABLE AND THE EXACT ROOT CAUSE CANNOT BE DETERMINED.

Additional Manufacturer Narrative · 0

THROUGH FOLLOW-UP COMMUNICATION LIVANOVA LEARNED THAT : THE DEVICE WAS CLEANED REGULARLY PER THE INSTRUCTION FOR USE THE HOSPITAL DOESN'T USE RE-USABLE BLANKETS PALL FILTER IS USED FOR TAP WATER THE DEVICE IS STORED EMPTY FOR LONG TERM STORAGE H2O2 IS ADDED WHEN THE WATER IN THE TANKS IS CHANGED (EACH 7 DAYS) HEATER-COOLER SURFACES AND WATER CIRCUITS ARE CLEANED AND DISINFECTED AS PER INSTRUCTION FOR USE IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 0

H.10: DOCUMENTATION OF MICROBIAL TESTS PERFORMED ON THIS SPECIFIC DEVICE WAS SHARED WITH LIVANOVA. ANALYSIS OF THE PROVIDED FILE REVEALED THAT THE UNIT TESTED POSITIVE FOR NTM SINCE JANUARY 24, 2022. IN ADDITION, IT WAS FOUND THAT THE DEVICE TESTED POSITIVE FOR NTM ALSO IN SEPTEMBER, NOVEMBER AND DECEMBER 2021. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 0

LIVANOVA (B)(4) RECEIVED A REPORT THAT A HEATER-COOLER SYSTEM 3T DEVICE WAS FOUND TO BE CONTAMINATED WITH NTM. THERE IS NO KNOWN PATIENT INVOLVEMENT.

Description of Event or Problem · 0

SEE INITIAL REPORT

Description of Event or Problem · 0

SEE INITIAL REPORT

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1079173 HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS DWC LIVANOVA DEUTSCHLAND 16-02-85

Patients

Seq Age Sex Outcome Treatment
1 Unknown