BD BBL¿ TRYPTICASE¿ SOY AGAR WITH 5% SHEEP BLOOD (TSA II)
Report
- Report Number
- 9680577-2022-00064
- Event Type
- Malfunction
- Date Received
- May 7, 2022
- Date of Event
- April 7, 2022
- Report Date
- July 12, 2022
- Manufacturer
- BECTON DICKINSON GMBH
- Product Code
- JSG
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
(B)(6). A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
H6 INVESTIGATION SUMMARY : THIS MEMO IS TO SUMMARIZE FINDINGS ON YOUR COMPLAINT (B)(4) AGAINST TRYPTICASE¿ SOY AGAR II WITH 5% SHEEP BLOOD, CATALOG NUMBER 254053, LOT NUMBER 2039481. EVENT DESCRIPTION: IT WAS REPORTED THAT THE PLATES WOULD HAVE DIFFERENT SHADES IN THE COLORATION OF THE MEDIA. FURTHERMORE, LEAKAGE AND CONTAMINATION WERE REPORTED. COMPLAINT HISTORY REVIEW: THE COMPLAINT TRENDS WERE REVIEWED, AND NO SIMILAR COMPLAINTS WERE IDENTIFIED FOR THIS BATCH NUMBER. THEREFORE, A TREND COULD NOT BE IDENTIFIED. SAMPLE ANALYSIS: PICTURE SAMPLE WAS PROVIDED FOR EVALUATION. BASED ON THE PICTURES, THE SURFACE OF THE MEDIA WAS FOUND TO BE UNEVEN, AND MOISTURE WAS DETECTED. THE SHADE OF THE MEDIUM IS DIFFERENT IN SOME PLATES. CONTAMINATION WAS NOT DETECTED. THE RETAIN SAMPLES WERE REVIEWED AND NO DEVIATION WAS DETECTED. EVALUATION SUMMARY AND INVESTIGATION CONCLUSION: BD MANUFACTURES VARIOUS PREPARED PLATED MEDIA TYPES, ALL OF WHICH CONTAIN AROUND 95 % WATER. MOISTURE, SURFACE ISSUES AND COLOR VARIATIONS WITHIN THE PLATES CAN BE CAUSED BY VARIOUS ENVIRONMENTAL FACTORS LINKED TO TRANSPORTATION AND STORAGE. SINCE THE RETAIN SAMPLES SHOWED NO DEVIATION, ALL QC RELEASE SPECIFICATIONS FOR THE PRODUCT WERE FULFILLED AND, ACCORDING TO THE COMPLAINT TRENDING, THIS INCIDENT CAN BE HANDLED AS ISOLATED ISSUE. NO DEVIATIONS COULD BE IDENTIFIED NEITHER DURING IN-PROCESS CONTROL NOR QUALITY CONTROL (QC) RELEASE. REVIEW OF THE COMPLAINT HISTORY SHOWED NO SIMILAR COMPLAINTS; THEREFORE, A TREND WAS NOT IDENTIFIED. BASED ON THE PICTURES PROVIDED, THIS COMPLAINT CAN BE CONFIRMED FOR COLOR VARIATION, UNEVEN SURFACE AND MOISTURE. A CONTAMINATION WAS NOT DETECTED AND THEREFORE, NOT CONFIRMED. SINCE ANY SYSTEMIC FAILURE CAN BE EXCLUDED FROM THE VALIDATED MANUFACTURING PROCESS, A DEFINITE ROOT CAUSE WAS NOT IDENTIFIED. BD WILL CONTINUE TO MONITOR INCOMING COMPLAINTS FOR SIMILAR DEFECT TYPES. H3 OTHER TEXT : NA.
IT WAS REPORTED THAT PRIOR TO USING BD BBL¿ TRYPTICASE¿ SOY AGAR WITH 5% SHEEP BLOOD (TSA II) THE CUSTOMER OBSERVED DISCOLORATION, LEAKING AGAR, AND BIOLOGICAL CONTAMINATION. THIS EVENT OCCURRED 15 TIME(S). THERE IS NO INDICATION THAT FALSE RESULTS WERE REPORTED TO HEALTH CARE PRACTITIONERS. THERE IS NO INDICATION OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "1. PLATES ALL HAD DIFFERENT COLORS (LIGHT RED, DARK RED). 2. BLOOD LEAKED FROM THE PLATES. 3. THE PLATES WERE PARTLY OVERGROWN WITH SOMETHING (CONTAMINATION). 4. SOME PLATES WERE TRANSPARENT. 5. PH VALUE WAS NOT RECORDED".
IT WAS REPORTED THAT PRIOR TO USING BD BBL¿ TRYPTICASE¿ SOY AGAR WITH 5% SHEEP BLOOD (TSA II) THE CUSTOMER OBSERVED DISCOLORATION, LEAKING AGAR, AND BIOLOGICAL CONTAMINATION. THIS EVENT OCCURRED 15 TIME(S). THERE IS NO INDICATION THAT FALSE RESULTS WERE REPORTED TO HEALTH CARE PRACTITIONERS. THERE IS NO INDICATION OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "1. PLATES ALL HAD DIFFERENT COLORS (LIGHT RED, DARK RED). 2. BLOOD LEAKED FROM THE PLATES. 3. THE PLATES WERE PARTLY OVERGROWN WITH SOMETHING (CONTAMINATION). 4. SOME PLATES WERE TRANSPARENT. 5. PH VALUE WAS NOT RECORDED."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1079133 | BD BBL¿ TRYPTICASE¿ SOY AGAR WITH 5% SHEEP BLOOD (TSA II) | CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL | JSG | BECTON DICKINSON GMBH | 2039481 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |