FDA Adverse Event Malfunction Summary report: N

RECIPROC FILES 4X

MDR report key: 14316403 · Received May 7, 2022

Report

Report Number
9611053-2022-00548
Event Type
Malfunction
Date Received
May 7, 2022
Report Date
June 30, 2022
Manufacturer
VDW GMBH
Product Code
EKS
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THERE HAS BEEN A PREVIOUS REPORT RECEIVED WITH A SIMILAR DEVICE WHERE THIS MALFUNCTION RESULTED IN A SERIOUS INJURY. THEREFORE, IT MUST BE PRESUMED THAT RECURRENCE OF THIS MALFUNCTION COULD POSSIBLY CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR REQUIRE MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE SUCH. AS SUCH, THIS EVENT IS REPORTABLE PER 21CFR PART 803. THE DEVICE IS AVAILABLE FOR EVALUATION, THOUGH HAS NOT BEEN RETURNED AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.

Additional Manufacturer Narrative · 0

INVOLVED PRODUCT THAT BROKE DURING USE WAS NOT RETURNED AND CANNOT BE ANALYZED. MOREOVER, NO UNUSED FILE IS AVAILABLE FOR EVALUATION. NOTHING UNUSUAL TO REPORT WAS FOUND DURING DHRS REVIEW (BATCHES #1689578, #1686275, #1686549, #1689858 AND #1689852). ROOT CAUSES ARE NOT IDENTIFIED. WE WILL TRACK THIS KIND OF EVENT AND MONITOR THE TREND.

Description of Event or Problem · 0

IN THIS EVENT IT IS REPORTED THAT A RECIPROC FILES 4X FILE BROKE DURING USE. BROKEN PART REMAINS IN CANAL AND BECAME PART OF FILLING. TREATMENT HAS BEEN COMPLETED. REQUEST FOR RETURN OF PRODUCT HAS BEEN MADE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2308782 RECIPROC FILES 4X FILE, PULP CANAL, ENDODONTIC EKS VDW GMBH 351627

Patients

Seq Age Sex Outcome Treatment