FDA Adverse Event Malfunction Summary report: N

CONSTELLATION SURGICAL PROCEDURE PAK

MDR report key: 14316290 · Received May 7, 2022

Report

Report Number
1644019-2022-00343
Event Type
Malfunction
Date Received
May 7, 2022
Date of Event
April 11, 2022
Report Date
August 5, 2022
Manufacturer
ALCON RESEARCH, LLC - HOUSTON
Product Code
LRO
PMA / PMN Number
K880961
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION PROVIDED IN D.9., H.3., H.6. AND H.10. A REVIEW OF THE DEVICE HISTORY RECORD TRACEABLE TO THE REPORTED LOT NUMBER INDICATES THAT THE PRODUCT WAS PROCESSED AND RELEASED ACCORDING TO THE PRODUCT¿S ACCEPTANCE CRITERIA. CURRENT TRACKING INDICATES NO ADVERSE TREND FOR THIS LOT FOR THIS EVENT. THE RETURNED SAMPLE WAS VISUALLY INSPECTED AND NO OBVIOUS DEFECTS WERE FOUND. NO OCCLUSION WAS OBSERVED IN THE MANIFOLDS. THE BALL IN THE CHECK VALVE OF THE DRIP CHAMBER MOVED FREELY PER SPECIFICATION. A CALIBRATED CONSTELLATION CONSOLE REPRESENTING THE CURRENT SOFTWARE VERSION WAS USED TO TEST THE SAMPLE. THE SAMPLE COULD PRIME AND TUNE WITH THE ULTRASONIC HANDPIECE SUCCESSFULLY. NO ANOMALIES WERE OBSERVED DURING PRIMING. NO SYSTEM MESSAGE CODE WAS GENERATED DURING TESTING. THE IRRIGATION PRESSURE RATE MET SPECIFICATION. FLUID FLOWED FROM THE BALANCED SALT SOLUTION (BSS) BOTTLE TO THE DRAIN BAG WITHOUT ANY INTERFERING. NO LEAKAGE WAS DETECTED FROM THE PUMP ELASTOMER OR ON THE PUMP AREA OF THE FLUIDICS MODULE. PROCESS WAS ABLE TO BE PERFORMED AFTER FUNCTIONAL TEST HAD COMPLETED. THE ROOT CAUSE OF THE CUSTOMER'S COMPLAINT COULD NOT BE ESTABLISHED; THE RETURNED SAMPLE FUNCTIONED PER SPECIFICATIONS. AFTER AN INVESTIGATION OF THIS COMPLAINT, IT HAS BEEN DETERMINED THAT THIS SAMPLE FUNCTIONED PER SPECIFICATIONS; THEREFORE, NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME. CURRENT TRACKING INDICATES NO ADVERSE TREND FOR THIS LOT FOR THIS EVENT AS THE PRODUCT MET SPECIFICATIONS. CURRENT TRACKING INDICATES NO ADVERSE TREND FOR THIS LOT FOR THIS EVENT AS THE PRODUCT MET SPECIFICATIONS. QUALITY ASSURANCE HAS REVIEWED THIS COMPLAINT AND WILL CONTINUE TO MONITOR DATA FOR EVIDENCE OF ADVERSE TRENDING AND TAKE FURTHER ACTION, AS APPROPRIATE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Description of Event or Problem · 0

A PHYSICIAN REPORTED THE BALANCED SALT SOLUTION DID NOT IRRIGATE DURING A CATARACT SURGERY. THE CASSETTE WAS REPLACED AND THE SURGERY WAS COMPLETED. THERE WAS NO PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1206003 CONSTELLATION SURGICAL PROCEDURE PAK GENERAL SURGERY TRAY (KIT) LRO ALCON RESEARCH, LLC - HOUSTON NA ASKU

Patients

Seq Age Sex Outcome Treatment
1 Unknown CONSTELLATION VISION SYSTEM