FDA Adverse Event Malfunction Summary report: N

PATROL PUMP

MDR report key: 1430528 · Received April 22, 2009

Report

Report Number
1527460-2009-00039
Event Type
Malfunction
Date Received
April 22, 2009
Date of Event
February 1, 2009
Report Date
March 10, 2009
Manufacturer
ABBOTT NUTRITION
Product Code
LZH
PMA / PMN Number
K943733
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

LZH PUMP, INFUSION, ENTERAL. THE TESTING AND INVESTIGATION RESULTS INDICATE THE COMPLAINT OF UNDER DELIVERY WAS CONFIRMED, DUE TO A STUTTERING MOTOR.

Description of Event or Problem · 1

THE REPORTER INDICATED THAT THE INTENDED RATE WAS 150ML/HR; HOWEVER, THE PUMP DELIVERED 45ML IN 30 MIN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PATROL PUMP FOR ENTERAL PUMPS LZH ABBOTT NUTRITION 52036

Patients

Seq Age Sex Outcome Treatment
1