FDA Adverse Event
Malfunction
Summary report: N
PATROL PUMP
MDR report key: 1430528
·
Received April 22, 2009
Report
- Report Number
- 1527460-2009-00039
- Event Type
- Malfunction
- Date Received
- April 22, 2009
- Date of Event
- February 1, 2009
- Report Date
- March 10, 2009
- Manufacturer
- ABBOTT NUTRITION
- Product Code
- LZH
- PMA / PMN Number
- K943733
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
LZH PUMP, INFUSION, ENTERAL. THE TESTING AND INVESTIGATION RESULTS INDICATE THE COMPLAINT OF UNDER DELIVERY WAS CONFIRMED, DUE TO A STUTTERING MOTOR.
Description of Event or Problem · 1
THE REPORTER INDICATED THAT THE INTENDED RATE WAS 150ML/HR; HOWEVER, THE PUMP DELIVERED 45ML IN 30 MIN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PATROL PUMP | FOR ENTERAL PUMPS | LZH | ABBOTT NUTRITION | 52036 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |