L-CATH PICC S/L 26GA (1.9F) 0.60MM X 30CM

MDR report key: 14304512 · Received May 6, 2022

Report

Report Number
0001625425-2022-01001
Event Type
Malfunction
Date Received
May 6, 2022
Date of Event
March 3, 2022
Report Date
June 21, 2022
Manufacturer
ARGON MEDICAL DEVICES
Product Code
LJS
UDI-DI
00886333209934
PMA / PMN Number
K091670
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE SAMPLE IS INDICATED AS RETURNED. AS OF THE DATE OF THIS REPORT, THE SAMPLE HAS NOT BEEN EVALUATED. A FOLLOW-UP REPORT WILL BE PROVIDED ONCE THE DEVICE HAS BEEN REVIEWED.

Additional Manufacturer Narrative · 0

A REVIEW OF THE DHR AND INSPECTION RECORDS WAS CONDUCTED, AND NO SIMILAR CONCERNS WERE FOUND. ONE CATHETER WAS RETURNED FOR REVIEW. VISUAL INSPECTION CONFIRMED A CRACK IN THE LUER. SEVERAL COMPLAINTS FOR THIS PART NUMBER HAVE PREVIOUSLY BEEN RECEIVED REGARDING A CRACKED HUB AND CAPA 2021-039 WAS INITIATED TO ADDRESS THIS ISSUE. THE CAPA IS CURRENTLY IN THE IMPLEMENTATION PHASE AND WILL EVALUATE THE CORRECTIVE ACTION IMPLEMENTATION FOR EFFECTIVENESS TO PREVENT A RECURRENCE OF THIS ISSUE. CAPA 2021-039 WAS INITIATED TO ADDRESS THIS ISSUE AND IS CURRENTLY IN THE IMPLEMENTATION PHASE. THE CAPA WILL EVALUATE THE CORRECTIVE ACTION IMPLEMENTATION FOR EFFECTIVENESS TO PREVENT A RECURRENCE OF THIS ISSUE.

Description of Event or Problem · 0

DURING INSERTION, CLINICIAN PLACED PICC, REMOVED THE STYLET, ATTACHED FLUSH, OBTAINED AIR, AND THE HUB CRACKED. BECAUSE THE PEEL-A-AWAY SHEATH WAS STILL INDWELLING IN THE PATIENT THIS PICC WAS REMOVED AND A NEW PICC INSERTED.

Description of Event or Problem · 0

DURING INSERTION, CLINICIAN PLACED PICC, REMOVED THE STYLET, ATTACHED FLUSH, OBTAINED AIR, AND THE HUB CRACKED. BECAUSE THE PEEL-A-AWAY SHEATH WAS STILL INDWELLING IN THE PATIENT THIS PICC WAS REMOVED AND A NEW PICC INSERTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2473213 L-CATH PICC S/L 26GA (1.9F) 0.60MM X 30CM L-CATH PICC LJS ARGON MEDICAL DEVICES 384539 11373440 00886333209934

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention| O