FDA Adverse Event Injury Summary report: N

QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 1

MDR report key: 14303478 · Received May 6, 2022

Report

Report Number
3005180920-2022-00328
Event Type
Injury
Date Received
May 6, 2022
Date of Event
April 6, 2022
Report Date
May 5, 2022
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JDI
UDI-DI
07630030802027
PMA / PMN Number
K082792
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 06 APRIL 2022: LOT 152958: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 07-OCT-2015. EXPIRATION DATE: 2020-09-12. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ALL ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT SINCE 2018. CLINICAL EVALUATION PERFORMED BY MEDACTA MEDICAL DIRECTOR: ABOUT 7 YEARS AFTER PRIMARY CEMENTLESS THA IN AN ELDERLY LADY, EXTENSIVE RADIOLUCENT LINES ARE VISIBLE ON THE RADIOGRPHIC CONTROL AROUND THE STEM. IN ABSENCE OF A SILENT INFECTION, THIS IS PROBABLY A CASE OF ASEPTIC LOOSENING, A POSSIBLE ADVERSE EVENT FOLLOWING THA, DESCRIBED IN LITERATURE. CAUSES REMAIN OFTEN UNDETERMINED. WE SEE NO DISCERNIBLE CLUES THAT COULD POINT TO A SPECIFIC CAUSE FOR THIS ADVERSE EVENT.

Description of Event or Problem · 0

REVISION SURGERY FOLLOWING STEM LOOSENING 6 YEARS AND 5 MONTHS AFTER PRIMARY. THE STEM, LINER, AND HEAD WERE REVISED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1205093 QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 1 HIP CEMENTLESS STEM JDI MEDACTA INTERNATIONAL SA 01.12.021 152958 07630030802027

Patients

Seq Age Sex Outcome Treatment
1 87 YR Female Required Intervention