QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 1
Report
- Report Number
- 3005180920-2022-00328
- Event Type
- Injury
- Date Received
- May 6, 2022
- Date of Event
- April 6, 2022
- Report Date
- May 5, 2022
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- JDI
- UDI-DI
- 07630030802027
- PMA / PMN Number
- K082792
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- 003
Narratives
BATCH REVIEW PERFORMED ON 06 APRIL 2022: LOT 152958: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 07-OCT-2015. EXPIRATION DATE: 2020-09-12. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ALL ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT SINCE 2018. CLINICAL EVALUATION PERFORMED BY MEDACTA MEDICAL DIRECTOR: ABOUT 7 YEARS AFTER PRIMARY CEMENTLESS THA IN AN ELDERLY LADY, EXTENSIVE RADIOLUCENT LINES ARE VISIBLE ON THE RADIOGRPHIC CONTROL AROUND THE STEM. IN ABSENCE OF A SILENT INFECTION, THIS IS PROBABLY A CASE OF ASEPTIC LOOSENING, A POSSIBLE ADVERSE EVENT FOLLOWING THA, DESCRIBED IN LITERATURE. CAUSES REMAIN OFTEN UNDETERMINED. WE SEE NO DISCERNIBLE CLUES THAT COULD POINT TO A SPECIFIC CAUSE FOR THIS ADVERSE EVENT.
REVISION SURGERY FOLLOWING STEM LOOSENING 6 YEARS AND 5 MONTHS AFTER PRIMARY. THE STEM, LINER, AND HEAD WERE REVISED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1205093 | QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 1 | HIP CEMENTLESS STEM | JDI | MEDACTA INTERNATIONAL SA | 01.12.021 | 152958 | 07630030802027 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 87 YR | Female | Required Intervention |