FDA Adverse Event Malfunction Summary report: N

UNKNOWN INSTERTER

MDR report key: 14291968 · Received May 5, 2022

Report

Report Number
0001825034-2022-01129
Event Type
Malfunction
Date Received
May 5, 2022
Date of Event
April 16, 2022
Report Date
May 12, 2022
Manufacturer
ZIMMER BIOMET, INC.
Product Code
JDH
PMA / PMN Number
N/A
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4). CUSTOMER HAS INDICATED THAT THE PRODUCT WILL NOT BE RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED: B4, B5, G3, H1, H2, H3, H6, H10 VISUAL EXAMINATION OF THE PROVIDED PICTURE IDENTIFIED THE TIP OF THE INSERTER IS FRACTURED. ADDITIONALLY THE DEVICE SHOWS SIGNS OF WEAR. COMPLAINT SAMPLE WAS EVALUATED AND THE REPORTED EVENT WAS CONFIRMED.. DEVICE HISTORY RECORD (DHR) REVIEW WAS UNABLE TO BE PERFORMED AS THE ITEM AND LOT NUMBER OF THE DEVICE INVOLVED IN THE EVENT IS UNKNOWN. ROOT CAUSE COULD NOT BE DETERMINED AS THE NECESSARY INFORMATION TO ADEQUATELY INVESTIGATE THE REPORTED EVENT WAS NOT PROVIDED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING AN INITIAL HIP ARTHROPLASTY, THE TIP OF THE THREADED INSERTER FRACTURED. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2473212 UNKNOWN INSTERTER PROSTHESIS, HIP JDH ZIMMER BIOMET, INC. N/A N/I

Patients

Seq Age Sex Outcome Treatment
1 Unknown