FDA Adverse Event Malfunction Summary report: N

PLATE MACCONKEY AGAR 100MM 10EA

MDR report key: 14289537 · Received May 5, 2022

Report

Report Number
9614033-2022-00035
Event Type
Malfunction
Date Received
May 5, 2022
Date of Event
April 7, 2022
Report Date
July 29, 2022
Manufacturer
BECTON DICKINSON DE MEXICO
Product Code
JSI
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THERE IS NO 510(K) FOR THIS DEVICE AS IT IS MANUFACTURED OUTSIDE THE US AND NOT SOLD IN THE US BUT IS CONSIDERED TO BE SUBSTANTIALLY SIMILAR TO THE LEGALLY U.S. MARKETED DEVICE BD BBL¿ MACCONKEY II AGAR CATALOG NUMBER 221172 WHICH IS A CLASS 1, EXEMPT DEVICE. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY: THE NOTIFIED BATCH FILE WAS REVIEWED, VERIFYING THAT ALL THE PROCESSES WERE FOLLOWED EFFECTIVELY DURING THE MANUFACTURE OF THE PRODUCT, LIKEWISE THE BATCH WAS ANALYZED BY THE QUALITY CONTROL DEPARTMENT AND THE RESULTS WERE SATISFACTORY FOR THE RELEASE OF THE BATCH. IT WAS NOT POSSIBLE TO CARRY OUT THE REVISION OF THE RETENTION SAMPLES BECAUSE THEY HAD ALREADY FULFILLED THEIR LIFE TIME, SO THEY WERE PURGED. IT IS CONCLUDED THAT THE EVENT REPORTED BY THE CLIENT IS CLASSIFIED AS CONFIRMED BASED ON THE PHOTOGRAPHIC EVIDENCE PROVIDED BY THE CLIENT, HOWEVER, A ROOT CAUSE ATTRIBUTABLE TO A FAILURE IN ANY OF THE PROCESS CONTROLS DURING THE PRODUCTION OR STORAGE PROCESS WAS NOT IDENTIFIED. WHILE THE PRODUCT WAS KEPT UNDER THE PROTECTION OF BD. THE PRODUCT ACCEPTS AN AQL OF 2.5, SO TWO CONTAMINATED PLATES IS ACCEPTABLE WITHIN THE SPECIFICATION CRITERIA. IT IS WORTH MENTIONING THAT THE PRODUCT WAS FOR 154 DAYS OUTSIDE THE BD SHELTER.

Description of Event or Problem · 0

IT WAS REPORTED THAT PRIOR TO USE WITH THE PLATE MACCONKEY AGAR 100MM 10EA, PLATES WERE CONTAMINATED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CLIENT SENDS THE FOLLOWING COMPLAINT ABOUT THE MAC CONKEY AGAR PLATE, LOT: 1288823, CA: (B)(6) 2022; DUE TO THE PRESENCE OF CONTAMINATION IN THEM.

Description of Event or Problem · 0

IT WAS REPORTED THAT PRIOR TO USE WITH THE PLATE MACCONKEY AGAR 100MM 10EA, PLATES WERE CONTAMINATED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CLIENT SENDS THE FOLLOWING COMPLAINT ABOUT THE MAC CONKEY AGAR PLATE, LOT: 1288823, CA: 05/12/22; DUE TO THE PRESENCE OF CONTAMINATION IN THEM."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1379863 PLATE MACCONKEY AGAR 100MM 10EA CULTURE MEDIA, SELECTIVE AND DIFFERENTIAL JSI BECTON DICKINSON DE MEXICO 1288823

Patients

Seq Age Sex Outcome Treatment
1 Unknown