FDA Adverse Event Malfunction Summary report: N

OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM

MDR report key: 14287859 · Received May 5, 2022

Report

Report Number
3004464228-2022-07350
Event Type
Malfunction
Date Received
May 5, 2022
Date of Event
April 30, 2022
Report Date
April 30, 2022
Manufacturer
INSULET CORPORATION
Product Code
LZG
UDI-DI
20385082000020
PMA / PMN Number
K192659
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ACCORDING TO THE COMPLAINANT THE DEVICE WILL NOT BE RETURNED FOR INVESTIGATION. WE ARE UNABLE TO CONFIRM THE REPORTED NEEDLE MECHANISM FAILURE OR TO DETERMINE ITS ROOT CAUSE. LOT RELEASE RECORDS WERE REVIEWED AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. SPECIFICALLY, A POD IS PAIRED TO A PDM AND PUT THROUGH SIMULATED USE TESTING INCLUDING COMMUNICATING WITH THE PDM, DEPLOYMENT, DELIVERING FLUID, OCCLUSION DETECTION, AND FREEDOM FROM HAZARD ALARMS. FOR: OMNI POD INSULIN MANAGEMENT SYSTEM ¿ USER GUIDE, MODEL: LZG, 18296-ENG-AW REV B. CHANGING YOUR POD CHAPTER 3 / PAGE 40. WARNINGS: DO NOT USE A POD IF IT IS PAST THE EXPIRATION DATE ON THE PACKAGE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE CANNULA DID NOT DEPLOY WHEN THE POD WAS ACTIVATED; THIS INDICATES A NEEDLE MECHANISM FAILURE OCCURRED. THE POD WAS NOT WORN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1678113 OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 18320 PD1C08301951 20385082000020

Patients

Seq Age Sex Outcome Treatment
1 14 YR Male