INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2022-07074
- Event Type
- Injury
- Date Received
- May 5, 2022
- Date of Event
- March 1, 2022
- Report Date
- March 14, 2023
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FTR
- UDI-DI
- 05060191607520
- PMA / PMN Number
- P020056
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.
PHOTO EVALUATION: VISUAL ANALYSIS OF THE PHOTOGRAPHS IDENTIFIED. LOT NUMBER IN THE PATCH: 3142262. OBSERVED BROKEN DEVICE. OBSERVED RED PARTICLES. DEVICE ANALYSIS PERFORMED THROUGH PHOTOGRAPHS, DUE TO THE IMPOSSIBILITY TO PERFORM MICROSCOPIC ANALYSIS IT IS NOT POSSIBLE TO DETERMINE THE MOST LIKELY FAILURE MODE. REFER TO PS-QP-ONEV-115717:COVID-19 QUALITY PLAN - COMPLAINT HANDLING.
DEVICE EVALUATION: BASED ON THE DEVICE ANALYSIS GRID, THE ASSESSMENTS OF THE COMPLAINT ARE: RUPTURE: OBSERVED, BROKEN DEVICE ON POSTERIOR SIDE ASSESSED AS UNIDENTIFIED (TEAR) OPENING (SHELL THICKNESS WITHIN SPECIFICATION). ADDITIONAL OBSERVATIONS: CREASE IS OBSERVED. WEAR ABRASION IS OBSERVED. MISSING SHELL ASSESSED AS INCONCLUSIVE. BLACK PARTICLES WERE OBSERVED ON THE SHELL.
PHYSICIAN REPORTS RIGHT SIDE RUPTURE. THE DEVICE HAS BEEN EXPLANTED.
PHYSICIAN REPORTS RIGHT SIDE RUPTURE. THE DEVICE HAS BEEN EXPLANTED.
HEALTHCARE PROFESSIONAL REPORTS RIGHT SIDE RUPTURE. THE DEVICE HAS BEEN EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1459522 | INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | ALLERGAN (COSTA RICA) | 3142262 | 05060191607520 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 49 YR | Female | Required Intervention |