FDA Adverse Event Injury Summary report: N

INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 14286316 · Received May 5, 2022

Report

Report Number
9617229-2022-07074
Event Type
Injury
Date Received
May 5, 2022
Date of Event
March 1, 2022
Report Date
March 14, 2023
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
UDI-DI
05060191607520
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.

Additional Manufacturer Narrative · 0

PHOTO EVALUATION: VISUAL ANALYSIS OF THE PHOTOGRAPHS IDENTIFIED. LOT NUMBER IN THE PATCH: 3142262. OBSERVED BROKEN DEVICE. OBSERVED RED PARTICLES. DEVICE ANALYSIS PERFORMED THROUGH PHOTOGRAPHS, DUE TO THE IMPOSSIBILITY TO PERFORM MICROSCOPIC ANALYSIS IT IS NOT POSSIBLE TO DETERMINE THE MOST LIKELY FAILURE MODE. REFER TO PS-QP-ONEV-115717:COVID-19 QUALITY PLAN - COMPLAINT HANDLING.

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: BASED ON THE DEVICE ANALYSIS GRID, THE ASSESSMENTS OF THE COMPLAINT ARE: RUPTURE: OBSERVED, BROKEN DEVICE ON POSTERIOR SIDE ASSESSED AS UNIDENTIFIED (TEAR) OPENING (SHELL THICKNESS WITHIN SPECIFICATION). ADDITIONAL OBSERVATIONS: CREASE IS OBSERVED. WEAR ABRASION IS OBSERVED. MISSING SHELL ASSESSED AS INCONCLUSIVE. BLACK PARTICLES WERE OBSERVED ON THE SHELL.

Description of Event or Problem · 0

PHYSICIAN REPORTS RIGHT SIDE RUPTURE. THE DEVICE HAS BEEN EXPLANTED.

Description of Event or Problem · 0

PHYSICIAN REPORTS RIGHT SIDE RUPTURE. THE DEVICE HAS BEEN EXPLANTED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTS RIGHT SIDE RUPTURE. THE DEVICE HAS BEEN EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1459522 INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 3142262 05060191607520

Patients

Seq Age Sex Outcome Treatment
1 49 YR Female Required Intervention