EXACTECH
Report
- Report Number
- 1038671-2022-00509
- Event Type
- Malfunction
- Date Received
- May 5, 2022
- Date of Event
- June 18, 2019
- Report Date
- May 4, 2022
- Manufacturer
- EXACTECH, INC.
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
AS A RESULT OF AN FDA INSPECTION CONDUCTED IN JAN 2020, EXACTECH, (B)(4), HAS COMMITTED TO REMEDIATING 3 YEARS OF COMPLAINTS (2017-2020). THIS MDR IS BEING SUBMITTED AS PART OF THAT REMEDIATION. FINDINGS: THE BUTTON COMPONENT (02-029-222-2005) OF THE TIBIAL TAMP GUIDE DISENGAGED FROM THE INSTRUMENT (02-029-22-2000). ENGINEERING PRINTS OF 02-029-122-2000 WERE REVIEWED. THE OVERLAP BETWEEN THE SNAP RING¿S OD AND THE ID OF THE HOUSING LIP IS .008 FOR THE WORST-CASE SCENARIO. ALTHOUGH THE OVERLAP IS POSITIVE, THE SMALL AMOUNT COULD CAUSE DISENGAGEMENT. MOST LIKELY UNDERLYING CAUSE/ROOT CAUSE: THE ROOT CAUSE FOR THE BUTTON DISENGAGEMENT, AS DETERMINED BY CAPA(B)(4), WAS LIKELY THE RESULT OF INSUFFICIENT OVERLAP BETWEEN THE SNAP RING AND THE LIP OF THE HOUSING TO PREVENT THE OCCURRENCE OF DISENGAGEMENT. CORRECTIVE ACTION TAKEN: PER (B)(4), THE COMMITTEE HAS DETERMINED THAT THE DESIGN WILL BE UPDATED, TESTED, AND IMPLEMENTED, AFTER WHICH A STAGED MARKET WITHDRAWAL WILL BE CONDUCTED. IN THE MEANTIME, THE EXACTECH TEAM WILL CONTINUE TO MONITOR THIS INSTRUMENT FOR THE ISSUE. CAPA(B)(4) IS STILL ONGOING AT THE TIME OF THIS EVALUATION, AND CORRECTIVE ACTIONS ARE SUBJECT TO CHANGE. IFU 700-096-181: INSTRUMENT INSPECTION: VISUALLY INSPECT THE INSTRUMENTS FOR DAMAGE SUCH AS FRACTURES; CRACKS; GOUGES; DEFORMATION; BURRS; DISCOLORATION, CORROSION, OR RUST; EXCESSIVE COMPONENT WEAR; NICKS ON CUTTING SURFACES, MISSING OR LOOSE COMPONENTS; BLOCKAGES IN CANNULAE, CLEANING HOLES OR OTHER CAVITIES THAT CANNOT BE REMOVED VIA STANDARD CLEANING; WORN OR DIFFICULT TO READ MARKINGS/ENGRAVINGS; OR OTHER APPARENT DAMAGE. CHECK THE FUNCTION OF MECHANISMS BY ACTUATING ANY LEVERS, KNOBS, SWITCHES, CONNECTORS, SLIDING FEATURES, HINGES, OR OTHER MECHANICAL INTERFACE FEATURES. ENSURE SMOOTH OPERATION OF THESE FEATURES OVER THEIR FUNCTIONAL RANGE OF MOTION. IF DAMAGE, WEAR, OR NON-FUNCTIONING/POORLY FUNCTIONING MECHANISMS ARE FOUND, DO NOT USE THE INSTRUMENT, AND CONTACT THE SALES REPRESENTATIVE OR CUSTOMER SERVICE FOR DISPOSITION. IFU STATES: THE SURGEON SHALL BECOME THOROUGHLY FAMILIAR WITH THE TECHNIQUE OF IMPLANTATION OF THE PROSTHESES BY: APPROPRIATE READING OF THE LITERATURE, AND TRAINING IN THE OPERATIVE SKILLS AND TECHNIQUES REQUIRED FOR SURGERY, AND REVIEWING INFORMATION REGARDING USE OF INSTRUMENTATION. IT IS A CLINICAL STANDARD OF PRACTICE IN THE OPERATING ROOM THAT ALL INSTRUMENTS SHOULD BE VISUALLY AND FUNCTIONALLY INSPECTED BEFORE USE, THESE GUIDELINES ARE FROM STANDARD ASSOCIATION OF PERI-OPERATIVE REGISTERED NURSES (AORN) GUIDELINES. SURGEONS ARE TO BE FAMILIAR AND KNOWLEDGEABLE WITH ALL INSTRUMENTATION, DEVICES AND PROFICIENT WITH SURGICAL TECHNIQUES. THIS DEVICE IS USED FOR TREATMENT NOT DIAGNOSIS. BASED ON REVIEW OF ALL AVAILABLE INFORMATION, THERE IS NO EVIDENCE TO REASONABLY SUGGEST THE REPORTED EVENT IS RELATED TO ANY MANUFACTURING ISSUES. THERE IS NO REPORTED PATIENT INVOLVEMENT. THE BUTTON CAN DISENGAGE DURING USE.
IT WAS REPORTED FROM THE US THAT THE TIBIA PUNCH TOWER 02-029-22-2000 FUNCTIONED DURING SURGERY BUT AFTER SPD CLEANING THE RELEASE BUTTON WAS FOUND MINUS THE SPRING. THE BUTTON CAN COME LOOSE AND POP OUT DURING USE. THE DEVICE WAS RECEIVED FOR ANALYSIS. THE REPRESENTATIVE WAS PRESENT AT THE TIME OF SURGERY. THERE IS NO ADDITIONAL INFORMATION FOR THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1564815 | EXACTECH | TRULIANT TIB TAMP GUIDE | LXH | EXACTECH, INC. | 82388001 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | Unknown |