FDA Adverse Event Malfunction Summary report: N

MINI ONE BALLOON BUTTON

MDR report key: 14279451 · Received May 4, 2022

Report

Report Number
1526012-2022-00006
Event Type
Malfunction
Date Received
May 4, 2022
Date of Event
February 26, 2022
Report Date
May 3, 2022
Manufacturer
APPLIED MEDICAL TECHNOLOGY, INC.
Product Code
KNT
PMA / PMN Number
K161413
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS A RESPONSE TO UF REPORT # (B)(4) WHICH WAS RECEIVED BY AMT ON 04/04/2022. BASED ON THE PROVIDED INFORMATION, THE COMPLAINT IS NOT A REPORTABLE EVENT PER 21 CFR SECTION 803. THERE WAS NO DEATH. THERE WAS NO SERIOUS INJURY, AS DEFINED BY THE FDA. THIS WAS NOT LIFE-THREATENING, NOR DID THIS RESULT IN PERMANENT IMPAIRMENT OR NECESSITATE MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE PERMANENT IMPAIRMENT. AMT WORKED WITH THE ORIGINAL REPORTER TO OBTAIN THE DEVICE FOR EXAMINATION AND PERFORMED AN INSPECTION OF THE DEVICE ON 04/26/2022. A TEAR IN THE BALLOON WALL WAS IDENTIFIED. A DEVICE HISTORY REVIEW WAS COMPLETED FOR THE REPORTED LOT NUMBER AND NO ANOMALIES WERE FOUND AND THERE HAVE BEEN NO OTHER COMPLAINTS REGARDING TEARS OR RUPTURES FROM THIS SAME BATCH. THE REPORTER INDICATED THAT THE DEVICE WAS IN USE FOR 130-DAYS PRIOR TO THE RUPTURE OF THE BALLOON. HOWEVER, IT WAS ALSO DETERMINED THAT THE LASERED LOT NUMBER ON THE RETURNED DEVICE WAS FROM A BATCH MANUFACTURED AFTER THE INDICATED 10/19/2021 PLACEMENT DATE. THEREFORE, EITHER THE REPORTED LONGEVITY IS INCORRECT, OR THE WRONG DEVICE WAS RETURNED. EITHER WAY, THE ACCURACY OF THE REPORTED INFORMATION IS IN QUESTION. THE COMPLAINT INFORMATION HAS BEEN LOGGED INTO OUR COMPLAINT DATABASE FOR TRENDING PURPOSES. COMPLAINT # (B)(4) WAS ASSIGNED TO THIS REPORT.

Description of Event or Problem · 0

PER THE ORIGINAL REPORTER IN UF/IMPORTER REPORT #: (B)(4), IT WAS REPORTED THAT "A VISITOR WAS HOLDING THE PATIENT WHEN SHE NOTICED A WET SPOT AND THEN, MOM SAW THAT THE G TUBE WAS OUT OF THE PATIENT. ONCE PATIENT WAS BACK TO BED, RN LOOKED AT THE TUBE AND THE BALLOON WAS CRACKED."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2473205 MINI ONE BALLOON BUTTON TUBES, GASTROINTESTINAL (AND ACCESSORIES) KNT APPLIED MEDICAL TECHNOLOGY, INC. M1-5-1215-I 210909-252

Patients

Seq Age Sex Outcome Treatment
1 Unknown