FDA Adverse Event Injury Summary report: N

WAGNER SL REVISION®, SCREW FOR TRIAL STEM, L 265

MDR report key: 14277784 · Received May 4, 2022

Report

Report Number
0009613350-2022-00267
Event Type
Injury
Date Received
May 4, 2022
Date of Event
March 16, 2022
Report Date
May 30, 2022
Manufacturer
ZIMMER SWITZERLAND MANUFACTURING GMBH
Product Code
JDI
UDI-DI
00889024277977
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MEDICAL PRODUCT: BIOLOX® DELTA, CERAMIC FEMORAL HEAD, L, 40/+3.5, TAPER 12/14 CATALOG#: 00-8775-040-03 ; LOT#: 2933958. WAGNER SL REVISION STEM 19/265 CATALOG#: 01.00102.619; LOT#: 3059866. G7 LONGEVITY NEUTRAL 40MM H CATALOG#:20104008 ; LOT#: 64994663. G7 PPS LTD ACET SHELL 64H CATALOG#: 01.0000.669 ; LOT#: 6129815. BONE SCR 6.5X35 SELF-TAP CATALOG#: 00625006535; LOT#: 65120781. BONE SCR 6.5X25 SELF-TAP CATALOG#: 00625006525; LOT#: J7051450. THE MANUFACTURER DID NOT RECEIVE THE DEVICE YET, HOWEVER IT IS INDICATED BY COMPLAINANT THAT IT WILL BE RETURNED FOR INVESTIGATION. THE LOT NUMBER OF THE DEVICE WAS RECEIVED. THE DEVICE HISTORY RECORDS WILL BE REVIEWED DURING INVESTIGATION. A CAUSE FOR THIS SPECIFIC EVENT CANNOT BE ASCERTAINED FROM THE INFORMATION PROVIDED. AS SOON AS SUPPLEMENTAL INFORMATION BECOMES AVAILABLE AN UPDATED REPORT WILL BE SUBMITTED. ZIMMER BIOMET¿S REFERENCE NUMBER OF THIS FILE IS (B)(4). IN TRANSIT TO MANUFACTURER.

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL AND/OR CORRECTED INFORMATION. UPDATED: D4 SERIAL NUMBER UNKNOWN. A CAUSE FOR THIS SPECIFIC EVENT CANNOT BE ASCERTAINED FROM THE INFORMATION PROVIDED. AS SOON AS SUPPLEMENTAL INFORMATION BECOMES AVAILABLE AN UPDATED REPORT WILL BE SUBMITTED. ZIMMER BIOMET¿S REFERENCE NUMBER OF THIS FILE IS (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT THE TRIAL LOCKING SCREW BROKE LEAVING THE DISTAL SEGMENT OF THE STEM IN THE FEMUR. AN ETO (EXTENDED TROCHANTERIC OSTEOTOMY) HAD TO BE PERFORMED TO EXTRACT THE DISTAL SEGMENT OF THE WAGNER SL STEM. PATIENT INVOLVEMENT - MORE COMPLEX SURGERY.

Description of Event or Problem · 0

NO CHANGE TO PREVIOUSLY REPORTED EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1218364 WAGNER SL REVISION®, SCREW FOR TRIAL STEM, L 265 HIP INSTRUMENT JDI ZIMMER SWITZERLAND MANUFACTURING GMBH N/A UNKNOWN 00889024277977

Patients

Seq Age Sex Outcome Treatment
1 Male Hospitalization