WAGNER SL REVISION®, SCREW FOR TRIAL STEM, L 265
Report
- Report Number
- 0009613350-2022-00267
- Event Type
- Injury
- Date Received
- May 4, 2022
- Date of Event
- March 16, 2022
- Report Date
- May 30, 2022
- Manufacturer
- ZIMMER SWITZERLAND MANUFACTURING GMBH
- Product Code
- JDI
- UDI-DI
- 00889024277977
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
MEDICAL PRODUCT: BIOLOX® DELTA, CERAMIC FEMORAL HEAD, L, 40/+3.5, TAPER 12/14 CATALOG#: 00-8775-040-03 ; LOT#: 2933958. WAGNER SL REVISION STEM 19/265 CATALOG#: 01.00102.619; LOT#: 3059866. G7 LONGEVITY NEUTRAL 40MM H CATALOG#:20104008 ; LOT#: 64994663. G7 PPS LTD ACET SHELL 64H CATALOG#: 01.0000.669 ; LOT#: 6129815. BONE SCR 6.5X35 SELF-TAP CATALOG#: 00625006535; LOT#: 65120781. BONE SCR 6.5X25 SELF-TAP CATALOG#: 00625006525; LOT#: J7051450. THE MANUFACTURER DID NOT RECEIVE THE DEVICE YET, HOWEVER IT IS INDICATED BY COMPLAINANT THAT IT WILL BE RETURNED FOR INVESTIGATION. THE LOT NUMBER OF THE DEVICE WAS RECEIVED. THE DEVICE HISTORY RECORDS WILL BE REVIEWED DURING INVESTIGATION. A CAUSE FOR THIS SPECIFIC EVENT CANNOT BE ASCERTAINED FROM THE INFORMATION PROVIDED. AS SOON AS SUPPLEMENTAL INFORMATION BECOMES AVAILABLE AN UPDATED REPORT WILL BE SUBMITTED. ZIMMER BIOMET¿S REFERENCE NUMBER OF THIS FILE IS (B)(4). IN TRANSIT TO MANUFACTURER.
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL AND/OR CORRECTED INFORMATION. UPDATED: D4 SERIAL NUMBER UNKNOWN. A CAUSE FOR THIS SPECIFIC EVENT CANNOT BE ASCERTAINED FROM THE INFORMATION PROVIDED. AS SOON AS SUPPLEMENTAL INFORMATION BECOMES AVAILABLE AN UPDATED REPORT WILL BE SUBMITTED. ZIMMER BIOMET¿S REFERENCE NUMBER OF THIS FILE IS (B)(4).
IT WAS REPORTED THAT THE TRIAL LOCKING SCREW BROKE LEAVING THE DISTAL SEGMENT OF THE STEM IN THE FEMUR. AN ETO (EXTENDED TROCHANTERIC OSTEOTOMY) HAD TO BE PERFORMED TO EXTRACT THE DISTAL SEGMENT OF THE WAGNER SL STEM. PATIENT INVOLVEMENT - MORE COMPLEX SURGERY.
NO CHANGE TO PREVIOUSLY REPORTED EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1218364 | WAGNER SL REVISION®, SCREW FOR TRIAL STEM, L 265 | HIP INSTRUMENT | JDI | ZIMMER SWITZERLAND MANUFACTURING GMBH | N/A | UNKNOWN | 00889024277977 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Male | Hospitalization |