FDA Adverse Event Malfunction Summary report: N

BINAXNOW COVID-19 AG CARD

MDR report key: 14274551 · Received May 4, 2022

Report

Report Number
1221359-2022-02286
Event Type
Malfunction
Date Received
May 4, 2022
Date of Event
March 15, 2022
Report Date
June 13, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
PMA / PMN Number
EUA202537
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. RELATED MFR REPORT NUMBERS: 1221359-2022-02281 THROUGH 1221359-2022-02289.

Additional Manufacturer Narrative · 0

INVESTIGATION REPORT: MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR KIT PART NUMBER 195-000/LOT 151701 AND DEVICE PART NUMBER 195-430H / LOT 146127A. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 151701 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD POSSIBLY BE RELATED TO THE SPECIFIC PATIENT SAMPLES.

Description of Event or Problem · 0

THE CUSTOMER REPORTED NINE (9) FALSE POSITIVE RESULTS WITH THE BINAXNOW COVID-19 AG CARD ON MULTIPLE DATES IN (B)(6) 2022 ON KITTED NASAL SWABS. THIS REPORT IS FOR PATIENT SIX (6) OF NINE (9) AND LOT 151701 (TOTAL QUANTITY 1). THE PATIENT WAS TESTED WITH THE BINAXNOW COVID-19 AG CARD ON (B)(6) 2022 WHICH GENERATED A POSITIVE RESULT. CONFIRMATION PCR TESTING WAS PERFORMED THE SAME DAY (PLATFORM:PERKINELMER) ON A NASAL WHICH GENERATED A NEGATIVE RESULT. THE PATIENT WAS SYMPTOMATIC AT THE TIME OF TESTING. NO SERIOUS INJURY, MEDICAL INTERVENTION OR TREATMENT DECISIONS OCCURRED AS A RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1218576 BINAXNOW COVID-19 AG CARD LATERAL FLOW IMMUNOASSAY IVD OF COVID-19 QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 151701

Patients

Seq Age Sex Outcome Treatment
1 34 YR Female