FDA Adverse Event Malfunction Summary report: N

L-CATH PICC S/L 28GA (1.2F) X 25CM 1 LUMEN

MDR report key: 14270454 · Received May 3, 2022

Report

Report Number
0001625425-2022-00999
Event Type
Malfunction
Date Received
May 3, 2022
Date of Event
March 27, 2022
Report Date
June 27, 2022
Manufacturer
ARGON MEDICAL DEVICES
Product Code
LJS
UDI-DI
00886333209910
PMA / PMN Number
K091670
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE SAMPLE IS INDICATED AS AVAILABLE FOR RETURN. AS OF THE DATE OF THIS REPORT, THE SAMPLE HAS NOT BEEN RETURNED. A FOLLOW-UP REPORT WILL BE PROVIDED ONCE THE DEVICE HAS BEEN RECEIVED AND REVIEWED.

Additional Manufacturer Narrative · 0

A REVIEW OF THE DHR AND INSPECTION RECORDS WAS CONDUCTED, AND NO SIMILAR CONCERNS WERE FOUND. ONE CATHETER WAS RETURNED FOR REVIEW. VISUAL INSPECTION CONFIRMED A CRACK IN THE LUER. SEVERAL COMPLAINTS FOR THIS PART NUMBER HAVE PREVIOUSLY BEEN RECEIVED REGARDING A CRACKED HUB AND CAPA 2021-039 WAS INITIATED TO ADDRESS THIS ISSUE. THE CAPA IS CURRENTLY IN THE IMPLEMENTATION PHASE AND WILL EVALUATE THE CORRECTIVE ACTION IMPLEMENTATION FOR EFFECTIVENESS TO PREVENT A RECURRENCE OF THIS ISSUE.

Description of Event or Problem · 0

¿PATIENT WAS TRANSFERRED FROM DOCTOR¿S MODESTO THAT HAD AN ARGON 1.2FR L-CATH PICC LINE IN PLACE. THE PICC HUB CRACKED DURING THE PATIENTS STAY WITH US.¿ PICC PLACED ¿ (B)(6) 2022. DATE OF EVENT ¿ (B)(6) 2022. PICC WAS D/C¿ED. NEW PICC PLACED.

Description of Event or Problem · 0

¿PATIENT WAS TRANSFERRED FROM DOCTOR¿S MODESTO THAT HAD AN ARGON 1.2FR L-CATH PICC LINE IN PLACE. THE PICC HUB CRACKED DURING THE PATIENTS STAY WITH US.¿ PICC PLACED ¿ (B)(6) 2022. DATE OF EVENT ¿ (B)(6) 2022. PICC WAS D/C¿ED. NEW PICC PLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2254274 L-CATH PICC S/L 28GA (1.2F) X 25CM 1 LUMEN L-CATH PICC LJS ARGON MEDICAL DEVICES 384516 11387617 00886333209910

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other