FDA Adverse Event Malfunction Summary report: N

BD BBL¿ TRYPTICASE SOY AGAR WITH 5% SHEEP BLOOD (TSA II)

MDR report key: 14270085 · Received May 3, 2022

Report

Report Number
1119779-2022-00640
Event Type
Malfunction
Date Received
May 3, 2022
Date of Event
April 7, 2022
Report Date
July 7, 2022
Manufacturer
BECTON, DICKINSON & CO. (SPARKS)
Product Code
JSG
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6 INVESTIGATION SUMMARY : THIS MEMO IS TO SUMMARIZE FINDINGS REGARDING THE COMPLAINT RELATED TO CATALOG NUMBER 221261, PLATE TRYPTICASE SOY AGAR 5% SB 100 EA, BATCH NUMBER 2040769 AND COMPLAINT NUMBER 4950369 FOR MISSING SLEEVE LABEL. DURING MANUFACTURING OF MATERIAL 221261, MEDIA IS FORMULATED AND SENT THROUGH A HIGH TEMPERATURE SHORT TIME STERILIZER TO REMOVE BIOBURDEN. THE PETRI DISHES ARE SUBJECTED TO UV RADIATION TO DECREASE BIOBURDEN. THE PETRI DISHES ARE FILLED IN A POSITIVE PRESSURE HEPA FILTERED ENVIRONMENT. THE FILLED PLATES ARE COOLED AND IMMEDIATELY WRAPPED INTO SLEEVES TO DECREASE THE INTRODUCTION OF MICROBES. SLEEVES ARE THEN PACKAGED INTO CARTONS AND THEN TRANSFERRED TO A REFRIGERATED TRUCK (2 TO 8 DEGREES C) FOR SHIPMENT TO THE DISTRIBUTOR. BD DISTRIBUTORS ARE PROVIDED WITH THE STORAGE GUIDELINES FOR THE SHIPPING AND HANDLING OF BD MEDIA OF 2 TO 8 DEGREES C IN A DARK PLACE. THE BATCH HISTORY RECORD FOR BATCH 2040769 WAS SATISFACTORY AND NO QUALITY NOTIFICATIONS WERE GENERATED DURING MANUFACTURING AND INSPECTION. THE RELEASE TESTING THAT IS PERFORMED ON THIS PRODUCT DOES INCLUDE PHYSICAL AND SLEEVE ATTRIBUTE TESTING. SAMPLE PLATES ARE INCUBATED AT 25 DEGREES C AND AT 35 DEGREES C FOR APPROXIMATELY 72 HOURS. THEY ARE TESTED FOR PHYSICAL AND SLEEVE ATTRIBUTES PRIOR TO RELEASE TO ENSURE THAT THEY CONFORM TO TYPICAL LEVELS. ALL PHYSICAL AND SLEEVE ATTRIBUTE TESTING PERFORMED ON THIS BATCH WAS SATISFACTORY PER OUR INTERNAL PROCEDURES. THE COMPLAINT HISTORY WAS REVIEWED, AND NO OTHER COMPLAINTS HAVE BEEN TAKEN ON BATCH 2040769. RETENTION SAMPLES FROM BATCH 2040769 WERE NOT AVAILABLE FOR INSPECTION. ONE UNOPENED SLEEVE (10 PLATES) FROM BATCH 2040769 WERE RETURNED IN A BOX WITH BUBBLE WRAP (TIME STAMP 0816). THE SLEEVE DID NOT HAVE A BD SLEEVE LABEL, HOWEVER THERE WAS A CUSTOMER APPLIED LABEL AND A REMNANT OF A TORN LABEL ON THE SLEEVE. THREE PHOTOS ALSO WERE RECEIVED FOR INVESTIGATION. THE PHOTOS EACH SHOW A SLEEVE WITH MEDIUM-RED AGAR PLATES AND WITH NO BD SLEEVE LABEL VISIBLE. THERE IS A CUSTOMER APPLIED LABEL AND THE REMNANT OF A TORN LABEL VISIBLE. THIS COMPLAINT CAN BE CONFIRMED. NO COMPLAINT TREND FOR MISSING SLEEVE LABELS HAS BEEN IDENTIFIED FOR THIS PRODUCT; NO ACTIONS ARE INDICATED AT THIS TIME. BD WILL CONTINUE TO TREND COMPLAINTS FOR MISSING SLEEVE LABELS.

Description of Event or Problem · 0

IT WAS REPORTED WHILE TESTING WITH BD BBL¿ TRYPTICASE SOY AGAR WITH 5% SHEEP BLOOD (TSA II) THE LABEL WAS MISSING FROM THE PRODUCT. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT THE SLEEVE MISSING THE PRODUCT LABEL. CALL ACTIVITY COMMENT: CUSTOMER PROBLEM: CUSTOMER REPORTS THAT SLEEVE MISSING THE PRODUCT LABEL FOR CAT 221261 LOT 2040769. INDICATION THAT CUSTOMER IS INDUSTRIAL INDU (IF APPLICABLE): INDU, IS A LETTER SIGNED BY QUALITY REQUIRED (Y/N)?: Y, REVIEWED SMAX CASE HISTORY: Y, HELP LIGHTNING ATTEMPT: N, HELP LIGHTNING SUCCESS: N. CUSTOMER REPORTS THAT SLEEVE MISSING THE PRODUCT LABEL FOR CAT 221261 LOT 2040769.

Description of Event or Problem · 0

IT WAS REPORTED WHILE TESTING WITH BD BBL¿ TRYPTICASE SOY AGAR WITH 5% SHEEP BLOOD (TSA II) THE LABEL WAS MISSING FROM THE PRODUCT. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT THE SLEEVE MISSING THE PRODUCT LABEL. CUSTOMER REPORTS THAT SLEEVE MISSING THE PRODUCT LABEL FOR CAT 221261 LOT 2040769.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2184195 BD BBL¿ TRYPTICASE SOY AGAR WITH 5% SHEEP BLOOD (TSA II) CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL JSG BECTON, DICKINSON & CO. (SPARKS) 2040769

Patients

Seq Age Sex Outcome Treatment
1 Unknown