FDA Adverse Event Injury Summary report: N

NV GXL LINER NEUTRAL, 36MM ID, GROUP 2 CUPS

MDR report key: 14269953 · Received May 3, 2022

Report

Report Number
1038671-2022-00504
Event Type
Injury
Date Received
May 3, 2022
Date of Event
April 1, 2022
Report Date
August 4, 2022
Manufacturer
EXACTECH, INC.
Product Code
JDI
UDI-DI
10885862207081
PMA / PMN Number
K121392
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

CONCOMITANT PRODUCTS: BIOLOX DELTA FEMORAL HEAD 36MM OD, +0MM (CAT# 170-36-00 / SERIAL# (B)(4)). ADDITIONAL INFORMATION, INCLUDING THE PRODUCT INVESTIGATION, WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT.

Additional Manufacturer Narrative · 0

AFTER FURTHER REVIEW OF ADDITIONAL INFORMATION. (H3) BASED ON THE AVAILABLE INFORMATION, THE PATIENT INVOLVED MEETS THE FOLLOWING RISK CRITERIA FOR EARLY PROSTHESIS WEAR/OSTEOLYSIS SIGNIFICANT EDGE LOADING OF THE FEMORAL HEAD IS CONFIRMED RADIOGRAPHICALLY AND ON THE PHOTOGRAPH OF THE ACETABULAR LINER, THE PATIENT IS NOTED TO HAVE UNDERGONE AN ANTERIOR APPROACH ON THE INDEX SURGERY, AND COMBINATION OF LARGEST AVAILABLE FEMORAL HEAD AND/OR THINNEST AVAILABLE ACETABULAR LINER WAS USED. THE MOST LIKELY CAUSE FOR THE REVISION REPORTED DUE TO EARLY PROSTHESIS WEAR IS A COMBINATION OF THE RISK FACTORS INCLUDING, USE ERROR, IMPLANT POSITIONING, IMPLANT SIZE SELECTION, AND PATIENT FACTORS (FITNESS FOR SURGERY, BIOMECHANICS, ACTIVITY LEVEL AND LOCAL TISSUE OXIDATION POTENTIAL COULD HAVE ALL CONTRIBUTED TO THE INCREASED TREND IN WEAR/OSTEOLYSIS RELATED COMPLAINTS). A RISK ASSESSMENT HAS BEEN INITIATED TO ESCALATE THIS ISSUE.

Description of Event or Problem · 0

AS REPORTED, APPROXIMATELY NINE YEARS POST INITIAL RIGHT THA, THIS 66 Y/O MALE PATIENT HAD A REVISION DUE TO SIGNIFICANT WEAR OF G2 36MM GXL NOVATION LINER. PATIENT WAS REVISED TO 36MM OPTION HEAD/+0 SLEEVE AND G2 36MM GXE NOVATION LINER. THERE WAS NO BREAKAGE OF DEVICE OR SURGICAL DELAY/PROLONGATION. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. THE DEVICE IS NOT AVAILABLE FOR EVALUATION DUE TO HSS KEEPS ALL RETRIEVALS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1762745 NV GXL LINER NEUTRAL, 36MM ID, GROUP 2 CUPS PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED JDI EXACTECH, INC. NV GXL LINER NEUTRAL, 36MM ID, GROUP 2 CUPS UNK 10885862207081

Patients

Seq Age Sex Outcome Treatment
1 66 YR Male Required Intervention - 6.5MM ACETABULAR BONE SCREW 25MM.| - ELEMENT-STEM, COLLARLESS W/HA, STD OFFSET, SZ 14.| - NV CROWN CUP CLSTR HOLE 52MM GROUP 2.