FDA Adverse Event Malfunction Summary report: N

CALCIUM

MDR report key: 14250500 · Received April 29, 2022

Report

Report Number
3002809144-2022-00145
Event Type
Malfunction
Date Received
April 29, 2022
Date of Event
April 10, 2022
Report Date
May 16, 2022
Manufacturer
ABBOTT GMBH
Product Code
CJY
UDI-DI
00380740161491
PMA / PMN Number
K062855
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

SID: 1000812: (B)(6). SID 10800712A: (B)(6). AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Additional Manufacturer Narrative · 0

THE COMPLAINT INVESTIGATION INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, SEARCH FOR SIMILAR COMPLAINTS, TICKET TRENDING REVIEW, LABELING REVIEW, AND DEVICE HISTORY REVIEW. RETURN TESTING WAS NOT COMPLETE AS RETURNS WERE NOT AVAILABLE. A SEARCH FOR SIMILAR TICKETS DID NOT REVEAL AN INCREASE IN COMPLAINT ACTIVITY. THE TREND REVIEW FOUND NO TRENDS RELATED TO THIS ISSUE. DEVICE HISTORY RECORD REVIEW DID NOT IDENTIFY ANY NON-CONFORMANCES OR DEVIATIONS RELATED TO THE CURRENT COMPLAINT ISSUE. LABELING WAS REVIEWED AND SUFFICIENTLY ADDRESSES THE CUSTOMER¿S ISSUE. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE CALCIUM REAGENT LOT 10782UN21 WAS IDENTIFIED.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED DISCREPANT CALCIUM RESULTS GENERATED ON AN ARCHITECT C4000 ANALYZER FOR TWO PATIENTS WHEN COMPARED TO RESULTS GENERATED ON ANOTHER ANALYZER. THE CUSTOMER PROVIDED REFERENCE RANGE IS 6.3-11.9 MG/DL. SID: (B)(4) INITIAL RESULT = 6.7 MG/DL, REPEAT =6.6 MG/DL, REPEAT ON ANOTHER ANALYZER = 6.0 MG/DL. SID (B)(4) INITIAL RESULT = 3.0 MG/DL, REPEAT RESULT = 3.1 MG/DL, REPEAT RESULT ON ANOTHER ANALYZER = 7.5 MG/DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
212764 CALCIUM AZO DYE, CALCIUM CJY ABBOTT GMBH 3L79-22 10782UN21 00380740161491

Patients

Seq Age Sex Outcome Treatment
1 Unknown ARC C4000 INTGR, 02P24-40, C400707.| ARC C4000 INTGR, 02P24-40, C400707.