CALCIUM
Report
- Report Number
- 3002809144-2022-00145
- Event Type
- Malfunction
- Date Received
- April 29, 2022
- Date of Event
- April 10, 2022
- Report Date
- May 16, 2022
- Manufacturer
- ABBOTT GMBH
- Product Code
- CJY
- UDI-DI
- 00380740161491
- PMA / PMN Number
- K062855
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
SID: 1000812: (B)(6). SID 10800712A: (B)(6). AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
THE COMPLAINT INVESTIGATION INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, SEARCH FOR SIMILAR COMPLAINTS, TICKET TRENDING REVIEW, LABELING REVIEW, AND DEVICE HISTORY REVIEW. RETURN TESTING WAS NOT COMPLETE AS RETURNS WERE NOT AVAILABLE. A SEARCH FOR SIMILAR TICKETS DID NOT REVEAL AN INCREASE IN COMPLAINT ACTIVITY. THE TREND REVIEW FOUND NO TRENDS RELATED TO THIS ISSUE. DEVICE HISTORY RECORD REVIEW DID NOT IDENTIFY ANY NON-CONFORMANCES OR DEVIATIONS RELATED TO THE CURRENT COMPLAINT ISSUE. LABELING WAS REVIEWED AND SUFFICIENTLY ADDRESSES THE CUSTOMER¿S ISSUE. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE CALCIUM REAGENT LOT 10782UN21 WAS IDENTIFIED.
THE CUSTOMER OBSERVED DISCREPANT CALCIUM RESULTS GENERATED ON AN ARCHITECT C4000 ANALYZER FOR TWO PATIENTS WHEN COMPARED TO RESULTS GENERATED ON ANOTHER ANALYZER. THE CUSTOMER PROVIDED REFERENCE RANGE IS 6.3-11.9 MG/DL. SID: (B)(4) INITIAL RESULT = 6.7 MG/DL, REPEAT =6.6 MG/DL, REPEAT ON ANOTHER ANALYZER = 6.0 MG/DL. SID (B)(4) INITIAL RESULT = 3.0 MG/DL, REPEAT RESULT = 3.1 MG/DL, REPEAT RESULT ON ANOTHER ANALYZER = 7.5 MG/DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 212764 | CALCIUM | AZO DYE, CALCIUM | CJY | ABBOTT GMBH | 3L79-22 | 10782UN21 | 00380740161491 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | ARC C4000 INTGR, 02P24-40, C400707.| ARC C4000 INTGR, 02P24-40, C400707. |