FDA Adverse Event Malfunction Summary report: N

TABLO

MDR report key: 14249830 · Received April 29, 2022

Report

Report Number
14249830
Event Type
Malfunction
Date Received
April 29, 2022
Date of Event
January 25, 2022
Report Date
April 22, 2022
Manufacturer
OUTSET MEDICAL, INC.
Product Code
KDI
Product Problem
Yes
Report Source
User Facility report
Reporter Location
MI
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Description of Event or Problem · 0

STOPPING THE CRRT RUN ON THE PT. AND PREPARING TO RETURN THE BLOOD TO THE PT. AND THE SCREEN TURNED BLACK AND AN UNKNOWN ALERT OF "UNFORTUNATELY, SEQUOIA GUI HAS STOPPED" AND THE MACHINE SHUTDOWN, REQUIRING THE NURSE TO MANUALLY RETURN THE BLOOD TO THE PATIENT. THE MACHINE WAS TAKEN DOWN TO THE STORE ROOM. THERE IT WAS PLUGGED IN AND CONNECTED TO THE WATER AND THE OUTPUT DRAIN. THE SCREEN ON THE MACHINE INDICATED THAT THE COMPANY SHOULD BE CALLED FOR SERVICE. I CALLED THE COMPANY AND NOTIFIED THE REPRESENTATIVE OF THE EVENT. HE ATTEMPTED TO REMOTE IN AND WAS UNABLE TO. HE WAS STILL ABLE TO RUN REPORTS. HE INSTRUCTED ME TO GO AHEAD AND RUN A HEAT CLEAN. I SET THE PUMP FOR A HEAT CLEAN. A CASE NUMBER OF WAS ASSIGNED AND I WAS INSTRUCTED TO TAG THE MACHINE AS OUT OF SERVICE. I THEN EMAILED ALL THE INDIVIDUALS THAT ARE ATTENDING THE DAILY TABLO TOUCHBASE MEETINGS." I WAS ON THE PHONE WITH THE REPRESENTATIVE FOR APPROXIMATELY 30 MINUTES WITH THE MACHINE. THIS DID NOT TAKE INTO ACCOUNT THE TIME IT TOOK TO GET THE MACHINE DOWN TO THE STORE ROOM, HOOK IT UP AND RUN THE HEAT CLEAN. IF I HAD NOT BEEN WORKING, THE CHARGE NURSE OR OTHER STAFF NURSE WOULD HAVE HAD TO DEAL WITH THIS AND BE OFF THE FLOOR FOR AN EXTENDED PERIOD OF TIME.

Description of Event or Problem · 0

STOPPING THE CRRT RUN ON THE PT. AND PREPARING TO RETURN THE BLOOD TO THE PT. AND THE SCREEN TURNED BLACK AND AN UNKNOWN ALERT OF "UNFORTUNATELY, SEQUOIA GUI HAS STOPPED" AND THE MACHINE SHUTDOWN, REQUIRING THE NURSE TO MANUALLY RETURN THE BLOOD TO THE PATIENT. THE MACHINE WAS TAKEN DOWN TO THE STORE ROOM . THERE IT WAS PLUGGED IN AND CONNECTED TO THE WATER AND THE OUTPUT DRAIN. THE SCREEN ON THE MACHINE INDICATED THAT THE COMPANY SHOULD BE CALLED FOR SERVICE. I CALLED THE COMPANY AND NOTIFIED THE REPRESENTATIVE OF THE EVENT. HE ATTEMPTED TO REMOTE IN AND WAS UNABLE TO. HE WAS STILL ABLE TO RUN REPORTS. HE INSTRUCTED ME TO GO AHEAD AND RUN A HEAT CLEAN. I SET THE PUMP FOR A HEAT CLEAN. A CASE NUMBER OF WAS ASSIGNED AND I WAS INSTRUCTED TO TAG THE MACHINE AS OUT OF SERVICE. I THEN EMAILED ALL THE INDIVIDUALS THAT ARE ATTENDING THE DAILY TABLO TOUCHBASE MEETINGS." I WAS ON THE PHONE WITH THE REPRESENTATIVE FOR APPROXIMATELY 30 MINUTES WITH THE MACHINE. THIS DID NOT TAKE INTO ACCOUNT THE TIME IT TOOK TO GET THE MACHINE DOWN TO THE STORE ROOM, HOOK IT UP AND RUN THE HEAT CLEAN. IF I HAD NOT BEEN WORKING, THE CHARGE NURSE OR OTHER STAFF NURSE WOULD HAVE HAD TO DEAL WITH THIS AND BE OFF THE FLOOR FOR AN EXTENDED PERIOD OF TIME.

Description of Event or Problem · 0

NO EQUIPMENT ISSUES OR HUMAN FACTORS INTERACTING WITH EQUIPMENT IDENTIFIED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
98750 TABLO DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM KDI OUTSET MEDICAL, INC.

Patients

Seq Age Sex Outcome Treatment
1 Unknown