FDA Adverse Event Malfunction Summary report: N

BD MICRO-FINE¿+ PEN NEEDLE

MDR report key: 14246805 · Received April 28, 2022

Report

Report Number
9616656-2022-00476
Event Type
Malfunction
Date Received
April 28, 2022
Date of Event
April 12, 2022
Report Date
May 20, 2022
Manufacturer
BECTON DICKINSON AND CO.
Product Code
FMI
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D9: DEVICE AVAILABLE FOR EVAL YES, D9: RETURNED TO MANUFACTURER ON: 10-MAY-2022. H6: INVESTIGATION SUMMARY: ONE OPEN 32G X 4MM PEN NEEDLE SAMPLE AND TWO PHOTOS WERE RETURNED FROM LOT. NO. 1160301, CAT. NO. 320136. A FUNCTIONALITY TEST AS PER Q-SOP-183-DL WAS CARRIED OUT ON THE RETURNED SAMPLE AND DAMAGE TO THE HUB WAS OBSERVED. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED AS PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. THIS ISSUE MOST LIKELY OCCURRED DUE TO MISALIGNMENT AT THE PICK AND PLACE STATION IN ASSEMBLY.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PEN NEEDLE WOULD NOT ATTACH TO THE PEN ON THE BD MICRO-FINE¿+ PEN NEEDLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: THE PEN NEEDLE COULD NOT BE ATTACHED TO THE PEN.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PEN NEEDLE WOULD NOT ATTACH TO THE PEN ON THE BD MICRO-FINE¿+ PEN NEEDLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM JAPANESE TO ENGLISH: THE PEN NEEDLE COULD NOT BE ATTACHED TO THE PEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
105560 BD MICRO-FINE¿+ PEN NEEDLE HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON DICKINSON AND CO. 1160301

Patients

Seq Age Sex Outcome Treatment
1 Unknown