HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP/DRIVELINE
Report
- Report Number
- 3007042319-2022-04904
- Event Type
- Injury
- Date Received
- April 28, 2022
- Date of Event
- May 29, 2021
- Report Date
- June 27, 2022
- Manufacturer
- HEARTWARE, INC.
- Product Code
- DSQ
- PMA / PMN Number
- P100047
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THIS INFORMATION IS BASED ENTIRELY ON JOURNAL LITERATURE. MEDTRONIC WAS MADE AWARE OF THIS EVENT THROUGH A SEARCH OF LITERATURE PUBLICATIONS. THE EVENT ONLY OCCURRED WITH ONE PATIENT BUT SPECIFIC DETAILS ON THE PATIENT WERE NOT PROVIDED. THE MODEL LISTED IN THE REPORT IS A REPRESENTATIVE OF THE MODEL FAMILY, AS THERE IS NO SPECIFIC MODEL LISTED. WITHOUT A LOT NUMBER OR DEVICE SERIAL NUMBER, THE MANUFACTURING DATE CANNOT BE DETERMINED. SINCE NO DEVICE ID WAS PROVIDED, IT IS UNKNOWN IF THIS EVENT HAS BEEN PREVIOUSLY REPORTED. A REQUEST FOR ADDITIONAL INFORMATION WILL BE MADE AND UPON RECEIPT A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY. REFERENCED ARTICLE: ASYMPTOMATIC VENTRICULAR FIBRILLATION IN CONTINUOUS FLOW LEFT-VENTRICULAR ASSIST DEVICE. AMERICAN JOURNAL OF EMERGENCY MEDICINE. 49 (2021) 130¿132. DOI:10.1016/J.AJEM.2021.05.065 INVESTIGATION OF THIS EVENT IS PENDING AND A SUPPLEMENTAL REPORT WILL BE SENT UPON ITS COMPLETION.
A SUPPLEMENTAL REPORT IS BEING SUBMITTED FOR INVESTIGATION COMPLETION. PRODUCT EVENT SUMMARY: ONE (1) PUMP WITH UNKNOWN SERIAL NUMBER WAS NOT RETURNED FOR EVALUATION. LOG FILE ANALYSIS COULD NOT BE PERFORMED SINCE LOG FILES COVERING THE REPORTED EVENT DATE WERE NOT AVAILABLE FOR ANALYSIS. AS A RESULT, THE REPORTED LOW FLOW EVENT COULD NOT BE CONFIRMED. BASED ON THE AVAILABLE INFORMATION, THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. BASED ON THE RISK DOCUMENTATION, POSSIBLE CAUSES OF THE LOW FLOW EVENT MAY BE ATTRIBUTED TO MULTIPLE FACTORS INCLUDING BUT NOT LIMITED TO THROMBUS AT THE INFLOW CANNULA/OUTFLOW GRAFT, CONSTRICTION AT THE OUTFLOW GRAFT, POOR VAD FILLING, INAPPROPRIATE PUMP ROTATIONAL SPEED. POSSIBLE CLINICAL FACTORS THAT MAY HAVE CONTRIBUTED TO THIS EVENT INCLUDE THE PATIENT¿S PRE-EXISTING HISTORY AND RELATED COMORBIDITIES, THE PROGRESSION OF THEIR UNDERLYING DISEASE, ISSUES RELATED TO THE THERAPEUTIC USE OF ANTICOAGULANT AND ANTIPLATELET MEDICATIONS AND THE PATIENT'S COMPLEX POST-OPERATIVE COURSE. THERE ARE POSSIBLE PATIENT, PHARMACOLOGICAL AND PROCEDURAL FACTORS THAT MAY HAVE CONTRIBUTED TO THIS EVENT. INVESTIGATION OF THIS EVENT IS COMPLETED, AND THE FILE WILL BE CLOSED. IF NEW INFORMATION IS RECEIVED, THE FILE WILL BE RE-OPENED, AND A SUPPLEMENTAL WILL BE SUBMITTED. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
A JOURNAL ARTICLE WAS REVIEWED THAT CONTAINED INFORMATION REGARDING VENTRICULAR FIBRILLATION (VF) IN CONTINUOUS FLOW LEFT VENTRICULAR ASSIST DEVICE (VAD). THE ARTICLE REPORTS A PATIENT WHO EXPERIENCED LIGHTHEADEDNESS AND THEIR VAD EXHIBITED LOW FLOWS WHICH TRIGGERED AN ALARM. A WEEK PRIOR, THEY HAD NAUSEA AND DIARRHEA. AFTER THE DEVICE ALARMED A SECOND TIME, THE PATIENT PRESENTED TO THE EMERGENCY DEPARTMENT AND WAS ASYMPTOMATIC. IN THE ED, A CARDIAC MONITOR NOTED THEM TO BE IN VF. THE PATIENT WAS SEDATED AND DEFIBRILLATED BUT REMAINED IN VF. THEY WERE DEFIBRILLATED A SECOND TIME WITH CONVERSION TO NORMAL SINUS RHYTHM CONDUCTION DELAY. SUBSEQUENTLY THE PATIENT WAS TREATED WITH INTRAVENOUS BETA-BLOCKERS AND A LOADING DOSE OF INTRAVENOUS AMIODARONE. THEY WERE THEN IMPLANTED WITH A CARDIOVERTER DEFIBRILLATOR (ICD) ON DAY TWO OF ADMISSION. THE STATUS/DISPOSITION OF THE VAD APPEARS TO BE STILL IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. FURTHER FOLLOW UP DID NOT YIELD ANY ADDITIONAL INFORMATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 214450 | HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP/DRIVELINE | VENTRICULAR (ASSISST) BYPASS | DSQ | HEARTWARE, INC. | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR | Female | Hospitalization| R |