FDA Adverse Event Malfunction Summary report: N

CADD EXTENSION SETS

MDR report key: 14246185 · Received April 28, 2022

Report

Report Number
3012307300-2022-07125
Event Type
Malfunction
Date Received
April 28, 2022
Report Date
May 5, 2023
Manufacturer
SMITHS MEDICAL ASD, INC.
Product Code
FPA
UDI-DI
15019517161953
PMA / PMN Number
K031361
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

NO PRODUCT OR PHOTOGRAPHIC EVIDENCE WERE PROVIDED TO AID IN THIS INVESTIGATION. THE COMPLAINT REPORT OFFERED INSUFFICIENT DETAILS TO DETERMINE WHETHER THIS PRODUCT FUNCTIONED AS INTENDED, OR WAS USED IN A MANNER CONSISTENT WITH ITS INSTRUCTIONS FOR USE (IFU) OR FAILED TO MEET PRODUCT SPECIFICATIONS. LACKING ANY ADDITIONAL EVIDENCE, THIS COMPLAINT HAS BEEN CLOSED AS UNCONFIRMED. IF THE PRODUCT IS RETURNED, SMITHS MEDICAL WILL REOPEN THIS COMPLAINT FOR FURTHER INVESTIGATION. A MANUFACTURING DHR REVIEW WAS NOT PERFORMED BECAUSE THE RESULTS OF THE COMPLAINT INVESTIGATION DO NOT INDICATE A PROBLEM WITH THE MANUFACTURE OF THE DEVICE.

Additional Manufacturer Narrative · 0

OTHER, OTHER TEXT: THIS MDR WAS GENERATED UNDER PROTOCOL (B)(4), AS A RESULT OF WARNING LETTER CMS# (B)(4). NO PROBLEMS OR ISSUES WERE IDENTIFIED DURING THIS DEVICE HISTORY RECORD REVIEW.

Description of Event or Problem · 0

(4/5, REFERENCE (B)(4) FOR RELATED COMPLAINTS) IT WAS REPORTED THAT UNDER- INFUSIONS OCCURRED WITH CADD PUMPS FOR (AMPICILLIN, CLOXICILLIN, PENICILLIN G). AT ITS BEST, IT WOULD UNDER-INFUSE BY AT LEAST 10% OF TOTAL MEDICATION VOLUME. AT THE HIGHEST, IT HAS UNDER-INFUSED BY NEARLY 50% OF TOTAL VOLUME. THIS HAS BEEN THE CASE FOR 5 DIFFERENT CLIENTS OF OURS. NO PATIENT INJURY WAS REPORTED.

Description of Event or Problem · 0

ADDITIONAL INFORMATION RECEIVED VIA EMAIL ON 09 MAY 2022: NO ADDITIONAL INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
110122 CADD EXTENSION SETS SET, ADMINISTRATION, INTRAVASCULAR FPA SMITHS MEDICAL ASD, INC. 21-7346-24 4235006 15019517161953

Patients

Seq Age Sex Outcome Treatment
1 Unknown