FDA Adverse Event Malfunction Summary report: N

BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE

MDR report key: 14245572 · Received April 28, 2022

Report

Report Number
3003152976-2022-00176
Event Type
Malfunction
Date Received
April 28, 2022
Date of Event
April 6, 2022
Report Date
May 24, 2022
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED AS A DEFAULT. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D9: DEVICE AVAILABLE FOR EVAL YES. D9: RETURNED TO MANUFACTURER ON: 04-MAY-2022. INVESTIGATION SUMMARY: SEVERAL SAMPLES WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THE PRODUCT WAS VISUALLY INSPECTED AND GREASE WAS FOUND ON ONE OF THE SYRINGES. A DEVICE HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT 2110092, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. FINAL PRODUCTS IN THIS MANUFACTURING LINE, FOR THIS REFERENCE ARE SAMPLED AND THEY ARE SUBJECTED TO VISUAL AND FUNCTIONAL INSPECTIONS DURING THE DIFFERENT MANUFACTURING SUB-PROCESSES ACCORDING TO PROCEDURES. MACHINES ROUTINELY UNDERGO PROPER MAINTENANCE AND CLEANING, QUALITY RECORDS ESTABLISH AS PROCESSES WERE CARRIED OUT NORMALLY. WHILE WE COULD NOT IDENTIFY A DIRECT ISSUE, THIS INCIDENT LIKELY OCCURRED DURING THE MOLDING PROCESS, THE GREASE ACCUMULATING WITHIN THE MOLD.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE HAD FOREIGN MATTER IN THE SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "DIRT FOUND INSIDE OF THE SYRINGE PACKAGING".

Description of Event or Problem · 0

IT WAS REPORTED THAT BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE HAD FOREIGN MATTER IN THE SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "DIRT FOUND INSIDE OF THE SYRINGE PACKAGING".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
115133 BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE PISTON SYRINGE FMF BECTON DICKINSON, S.A. 2110092

Patients

Seq Age Sex Outcome Treatment
1 Unknown