FDA Adverse Event Other Summary report: N

TRACH CARE NEONATAL "Y" 6F

MDR report key: 1424218 · Received March 27, 2009

Report

Report Number
8030647-2009-00006
Event Type
Other
Date Received
March 27, 2009
Date of Event
September 8, 2008
Report Date
March 25, 2009
Manufacturer
AVENT S. DE R.L. DE C.V.
Product Code
BSY
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CATHETER TIP WAS RECEIVED AND EVALUATED. THE TIP SHOWED A DIAGONAL CUT. BASED UPON THE EVAL OF THE INCIDENT SAMPLE AND INVESTIGATIONS FOR SIMILAR COMPLAINTS FOR THE FAILURE MODE IDENTIFIED, THIS COMPLAINT IS RELATED TO A USER ERROR "CUT CATHETER". IT SHOULD BE NOTED THAT IN THE PRODUCT'S DIRECTIONS FOR USE, THE WARNING LABEL CLEARLY STATES "FULLY WITHDRAW TRACH CARE CATHETER BEFORE CUTTING ENDOTRACHEAL TUBE, OTHERWISE THE CATHETER MAY ALSO BE CUT." NO ADDITIONAL INFORMATION HAS BEEN RECEIVED. WE WILL CONTINUE TO CLOSELY MONITOR THE FIELD PERFORMANCE OF THIS PRODUCT TO IDENTIFY EMERGING TRENDS AND IF APPLICABLE, ADDITIONAL INVESTIGATIONS WILL BE INITIATED AND CORRECTIVE ACTION TAKEN IF INDICATED. INFO FROM THIS INCIDENT WILL BE INCLUDED IN OUR PRODUCT COMPLAINT & MDR TREND REPORTING SYSTEMS. ONGOING ANALYSIS OF TREND INFO IS USED TO IDENTIFY THE NEED FOR ADDITIONAL INVESTIGATIONS.

Description of Event or Problem · 1

IT HAS BEEN REPORTED TO KIMBERLY-CLARK THAT WHILE THE STAFF WAS ATTEMPTING TO BOTTLE FEED AN INFANT, THE INFANT VOMITED A 2" PLASTIC TUBE WITH A BLACK TIP. THIS EVENT TOOK PLACE IN (B)(6) 2008 AND WAS NOT REPORTED TO THE KIMBERLY CLARK SALES REP UNTIL (B)(6), 2008. THERE WAS NO REPORT OF ANY SERIOUS INJURIES TO THE INFANT. THE INFANT IS REPORTED TO BE IN STABLE CONDITION. KIMBERLY-CLARK HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFO THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE KIMBERLY-CLARK COMPLAINT DATABASE AND IDENTIFIED AS (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRACH CARE NEONATAL "Y" 6F TRACH CARE BSY AVENT S. DE R.L. DE C.V. 426867

Patients

Seq Age Sex Outcome Treatment
1