ID NOW COVID-19 ASSAY
Report
- Report Number
- 1221359-2022-02138
- Event Type
- Malfunction
- Date Received
- April 28, 2022
- Date of Event
- April 2, 2022
- Report Date
- June 3, 2022
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC
- Product Code
- QJR
- UDI-DI
- 10811877011269
- PMA / PMN Number
- EUA2000074
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE INVESTIGATION IS STILL IN PROGRESS . A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. REFERENCE MFR. REPORTS:1221359-2022-02136 THROUGH 1221359-2022-02141.
TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 1054629 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-000 / LOT 1054629 , TEST BASE PART NUMBER 190-430 / LOT 1054629. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, A POSSIBLE ASSIGNABLE ROOT CAUSE IS PATIENT SAMPLE INTERFERENCE. REFERENCE MFR. REPORTS:1221359-2022-02136 THROUGH 1221359-2022-02141.
THE CUSTOMER REPORTED SIX (6) FALSE POSITIVE RESULTS WITH THE IDNOW COVID-19 ASSAY PERFORMED ON (B)(6) 2022 ON MIDTURBINATE NASAL .CONFORMATION TEST (PCR), WAS PERFORMED ON (B)(6) 2022,NASOPHARYNGEAL SWABS ,AND GENERATED NEGATIVE RESULT. ADDITIONALLY, THE CUSTOMER CONFIRMED THERE WAS A DELAY ON STAFF TESTED POSITIVE,MISSED WORK, WERE FOUND NEGATIVE THROUGH PCR TESTING.NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME,WAS PROVIDED. THIS REPORT IS THREE (3) OF SIX (6) AND ADDRESSES LOT (1054629) TOTAL QUANTITY (2).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 58508 | ID NOW COVID-19 ASSAY | MOLECULAR IVD FOR ID NOW COVID-19 | QJR | ABBOTT DIAGNOSTICS SCARBOROUGH, INC | 1054629 | 10811877011269 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 31 YR | Female |