FDA Adverse Event Other Summary report: N

PHASEAL INFUSION ADAPTER

MDR report key: 1420442 · Received May 21, 2009

Report

Report Number
9615998-2009-00008
Event Type
Other
Date Received
May 21, 2009
Date of Event
December 1, 2008
Report Date
May 19, 2009
Manufacturer
CARMEL PHARMA AB.
Product Code
LHI
PMA / PMN Number
K023747
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PHASEAL INFUSION ADAPTER C100 IS DESIGNED IN ACCORDANCE WITH THE REQUIREMENTS OF INTERNATIONAL STANDARDS, (B) (4). THE HOLDING FORCE BETWEEN THE PHASEAL INFUSION ADAPTER C100 AND PORTS OF VARIOUS INFUSION CONTAINERS HAS BEEN TESTED AND FOUND TO BE IN ACCORDANCE WITH STANDARD REQUIREMENTS.

Description of Event or Problem · 1

REPORT OF INFUSION ADAPTER FALLING OUT OF INFUSION BAG PORT. CHEMO SPILLED AND A SMALL AMOUNT WAS SPLASHED ONTO PHARMACY TECH'S FACE AND EYES. HOSPITAL PROTOCOL WAS FOLLOWED AND THERE ARE NO RESIDUAL EFFECTS OF THE SPILL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PHASEAL INFUSION ADAPTER INTRAVASCULAR ADMINISTRATION SET LHI CARMEL PHARMA AB. C100 8420439

Patients

Seq Age Sex Outcome Treatment
1