FDA Adverse Event Malfunction Summary report: N

BD PRESET¿ ARTERIAL BLOOD COLLECTION SYRINGE

MDR report key: 14197631 · Received April 25, 2022

Report

Report Number
9617032-2022-00419
Event Type
Malfunction
Date Received
April 25, 2022
Date of Event
March 20, 2022
Report Date
April 25, 2022
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
JKA
PMA / PMN Number
K022426
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: "MATERIAL #: 364314, LOT/BATCH #: 1162363. BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 10 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY FUNCTIONAL TESTING, EACH USED TO DRAW WATER, AND NO ISSUES WERE OBSERVED RELATING TO DIFFICULT TO MOVE PLUNGER OR LEAKAGE AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODES DIFFICULT TO MOVE PLUNGER AND LEAKAGE. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODES. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS."

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD PRESET¿ ARTERIAL BLOOD COLLECTION SYRINGE, BLOOD FLOW REACHES 0.5ML, THE PISTON OF THE ARTERIAL BLOOD COLLECTION DEVICE CANNOT SLIDE BY ITSELF, SO 1ML BLOOD CAN BE EXTRACTED MANUALLY, AND BLOOD LEAKAGE OCCURRED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: ¿WHEN THE ARTERIAL BLOOD COLLECTION DEVICE IS CONNECTED TO THE BLOOD COLLECTION NEEDLE, BLOOD IS DRAWN SMOOTHLY. WHEN THE ARTERIAL BLOOD FLOW REACHES 0.5ML, THE PISTON OF THE ARTERIAL BLOOD COLLECTION DEVICE CANNOT SLIDE BY ITSELF, SO 1ML BLOOD CAN BE EXTRACTED MANUALLY, AND BLOOD LEAKAGE OCCURRED."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
969952 BD PRESET¿ ARTERIAL BLOOD COLLECTION SYRINGE BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON AND COMPANY (BD) 1162363

Patients

Seq Age Sex Outcome Treatment
1 Unknown