FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19 ASSAY

MDR report key: 14196479 · Received April 25, 2022

Report

Report Number
1221359-2022-02297
Event Type
Malfunction
Date Received
April 25, 2022
Date of Event
March 3, 2022
Report Date
May 19, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
UDI-DI
10811877011320
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PMA/510K: SIMILAR TO 190-000. THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.

Additional Manufacturer Narrative · 0

INVESTIGATION REPORT: TESTING WAS PERFORMED IN DUPLICATE AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 1060760 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FO RKIT PART NUMBER 191-000 / LOT 1060760 AND TEST BASE PART NUMBER 190-430 / LOT 1060760. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE NEGATIVE AND FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 1060760 SHOWED THAT THE COMPLAINT RATE IS (B)(4). IN CONCLUSION, THE RETENTION TESTING YIELDED EXPECTED RESULTS WHEN TESTING INTERNAL QC SAMPLES. THE MANUFACTURING BATCH RECORD REVIEW REVEALED THAT THE PRODUCT MET ACCEPTANCE CRITERIA FOR RELEASE. ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, A POSSIBLE ASSIGNABLE ROOT CAUSE IS PATIENT SAMPLE INTERFERENCE. SUBSTANCES IN THE SAMPLE ITSELF MAY HAVE INTERFERED WITH THE REACTION. REVIEW OF COMPLAINTS AGAINST THE KIT LOT(S) FOR THE REPORTED ISSUE INDICATES PRODUCT PERFORMANCE IS AS EXPECTED AND HAVE NOT IDENTIFIED A PRODUCT DEFICIENCY.

Description of Event or Problem · 0

THE CUSTOMER REPORTED FALSE POSITIVE RESULTS WITH THE ID NOW COVID-19 ASSAY PERFORMED ON (B)(6) 2022. CONFIRMATION PCR TESTING WAS PERFORMED ON AN UNKNOWN DATE WHICH GENERATED A NEGATIVE RESULT (PLATFORM AND SAMPLE TYPE UNKNOWN). NO FURTHER INFORMATION REGARDING SYMPTOMS, HEALTH IMPACT OR OUTCOME WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
690478 ID NOW COVID-19 ASSAY MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 1060760 10811877011320

Patients

Seq Age Sex Outcome Treatment
1 43 YR Male