FDA Adverse Event Injury Summary report: N

UNKNOWN ENDO GIA SULU

MDR report key: 14194509 · Received April 25, 2022

Report

Report Number
1219930-2022-01469
Event Type
Injury
Date Received
April 25, 2022
Date of Event
December 4, 2021
Report Date
April 25, 2022
Manufacturer
COVIDIEN LP LLC NORTH HAVEN
Product Code
GDW
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

TITLE: GASTRIC REMNANT MUCOCELE AFTER CONVERSION OF VERTICAL BAND GASTROPLASTY TO ONE ANASTOMOSIS GASTRIC BYPASS: A VIDEO VIGNETTE SOURCE: OBESITY SURGERY (2022) 32:955¿956, RECEIVED: 14 SEPTEMBER 2021 / REVISED: 9 NOVEMBER 2021 / ACCEPTED: (B)(6) 2021 © THE AUTHOR(S), UNDER EXCLUSIVE LICENCE TO SPRINGER SCIENCE+BUSINESS MEDIA, LLC, PART OF SPRINGER NATURE 2021. HTTPS://DOI.ORG/10.1007/S11695-021-05842-8. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

ACCORDING TO LITERATURE SOURCE OF STUDY PERFORMED ON A (B)(6) FEMALE WITH A HISTORY OF VENOUS BLOOD GAS (VBG) FOR MORBID OBESITY, TWO (2) YEARS BEFORE AND BACKGROUND OF HYPERTENSION AND HYPERLIPIDEMIA, THE PATIENT UNDERWENT LAPAROSCOPIC ONE ANASTOMOSIS GASTRIC BYPASS CONVERSION SURGERY AFTER VERTICAL BANDED GASTROPLASTY. THE GASTRIC POUCH WAS CREATED AND GASTROJEJUNOSTOMY WAS PERFORMED USING A PURPLE RELOAD. THREE WEEKS LATER THE PATIENT HAD A LEFT-UPPER QUADRANT ABDOMINAL PAIN. ABDOMINAL COMPUTERIZED TOMOGRAPHY (CT) SCAN REVEALED A HUGE LEFT SUB DIAPHRAGMATIC COLLECTION ABOUT FOURTEEN CENTIMETERS. THE PATIENT WAS TREATED WITH A PIGTAIL DRAIN INSERTED UNDER ULTRASOUND GUIDANCE. THE DRAIN WAS REMOVED AFTER 2 WEEKS. ONE MONTH LATER, THE FLUID COLLECTION REOCCURRED AND THE PATIENT UNDERWENT REOPERATION WHICH REVEALED A MUCOCELE OF THE GASTRIC REMNANT WHICH WAS RESECTED. IT WAS NOTED THAT MUCOCELE OF THE GASTRIC REMNANT CAN OCCUR WHEN THE STAPLER WAS FIRED TOO CLOSE TO PREVIOUS STAPLER LINE. THE PATIENT HAD AN UNEVENTFUL RECOVERY AND DISCHARGED FORTY EIGHT HOURS POSTOPERATIVELY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
195067 UNKNOWN ENDO GIA SULU STAPLE, IMPLANTABLE GDW COVIDIEN LP LLC NORTH HAVEN UNKNOWN ENDO GIA SULU

Patients

Seq Age Sex Outcome Treatment
1 45 YR Female Hospitalization| R