FDA Adverse Event Other Summary report: N

Q-SYTE LUER ACCESS SITE

MDR report key: 1419343 · Received July 20, 2009

Report

Report Number
9610847-2009-00043
Event Type
Other
Date Received
July 20, 2009
Report Date
June 22, 2009
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS
Product Code
FPA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

RESULTS - RECEIVED ONE USED SAMPLE FOR THE INVESTIGATION. OUR QUALITY ENGINEER VISUALLY AND MICROSCOPICALLY INSPECTED THE RETURNED SAMPLE AND OBSERVED TOP SEPTUM DAMAGE AND AIR BUBBLES. THE DEVICE HISTORY COULD NOT BE REVIEWED AS THE LOT NUMBER IS UNKNOWN. CONCLUSIONS - THE ENGINEER CONCLUDED THAT THE DAMAGE TO THE TOP SEPTUM MAY HAVE BEEN CAUSED BY ACTUATIONS GREATER THAN 100 AND/OR A SHARP MALE LUER INSERTED INTO THE BD Q-SYTE. A ROOT CAUSE COULD NOT BE DETERMINED SINCE THERE WAS NO INFORMATION ABOUT THE NUMBER OF ACTUATIONS PERFORMED AND TYPE OF MALE LUER USED BY THE CUSTOMER. THE INSTRUCTIONS FOR USE STATE: FOR PROPER USE, CLINICIANS MUST BE FAMILIAR WITH AND TRAINED IN THE USE OF THE BD Q-SYTE DEVICE. THE BD Q-SYTE DEVICE IS COMPATIBLE FOR USE ONLY WITH ISO LUER SLIP AND LUER LOCK CONNECTORS PROVIDED ON STANDARD IV ADMINISTRATION SETS, EXTENSION SETS, AND SYRINGES. FIRMLY AND COMPLETELY PUSH THE SYRINGE OR ADMINISTRATION SET INTO THE ACCESS PORT UNTIL SECURE CONNECTION HAS BEEN MADE. WHEN CONNECTING TO OR DISCONNECTING FROM THE BD Q-SYTE DEVICE, ALWAYS INSERT OR REMOVE THE MALE LUER USING A "STRAIGHT -ON/STRAIGHT-OFF APPROACH. ANGLED INSERTIONS/REMOVAL MAY CAUSE POOR FUNCTIONING AND/OR DAMAGE TO THE DEVICE. DO NOT EXCEED 100 ACTUATIONS OF THE DEVICE. (B)(4).

Description of Event or Problem · 1

WHILE WITHDRAWING ON THE PROXIMAL STOPCOCK OF THE CATHETER, THE NURSE DETECTED AIR BUBBLES. THE SEPTUM OF THE BD Q-SYTE WAS SLIGHTLY OPENED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 Q-SYTE LUER ACCESS SITE INTRAVASCULAR ADMINISTRATION SET FPA BECTON DICKINSON INFUSION THERAPY SYSTEMS NA UNK

Patients

Seq Age Sex Outcome Treatment
1 UNK Other