PLATE COLUMBIA AG 5PRCT SB 90MM
Report
- Report Number
- 9680577-2022-00061
- Event Type
- Malfunction
- Date Received
- April 22, 2022
- Date of Event
- March 9, 2022
- Report Date
- July 5, 2022
- Manufacturer
- BECTON DICKINSON GMBH
- Product Code
- KZI
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NO
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
H.10 MANUFACTURE NARRATIVE: COMPLAINT HISTORY REVIEW: COMPLAINT HISTORY WAS REVIEWED AND A TREND WAS IDENTIFIED FOR THIS PRODUCT. BATCH HISTORY RECORD (BHR) REVIEW: THE BHR WAS REVIEWED. NO DEVIATIONS FROM VALIDATED MANUFACTURING PROCESSES WERE IDENTIFIED. SAMPLE ANALYSIS: RETAIN SAMPLES WERE REVIEWED AND INCUBATED. NO CONTAMINATION OBSERVED AFTER THE INCUBATION. RETURN SAMPLES WERE NOT PROVIDED AND THE CONTAMINATION CANNOT BE IDENTIFIED ON THE SHARED PICTURE. EVALUATION RESULTS: THIS PRODUCT IS FILLED UNDER ASEPTIC CONDITIONS. THEREFORE STERILITY OF THE FINISHED PRODUCT CANNOT BE GUARANTEED. UNFORTUNATELY A 100 % INSPECTION LEVEL FOR STERILITY IS NOT POSSIBLE AND STERILITY TESTING IS CARRIED OUT ON THE BASIS OF A REPRESENTATIVE SAMPLE. IN CONSEQUENCE, OCCASIONAL CONTAMINATIONS CANNOT BE PREVENTED. IN THIS CASE, A TREND WAS IDENTIFIED. CURRENTLY SEVERAL CROSS-FUNCTIONAL TEAMS ARE CONDUCTING EXTENSIVE INVESTIGATION TO REDUCE THE OCCURRENCE FOR SUCH SUBLIMINAL CONTAMINATIONS. PLEASE BE AWARE THAT SUCH INVESTIGATION WILL TAKE ITS TIME. A DEFINITE ROOT CAUSE WAS NOT IDENTIFIED. BD WILL CONTINUE TO MONITOR INCOMING COMPLAINTS FOR SIMILAR DEFECT TYPES. INVESTIGATION CONCLUSION: ALTHOUGH CONTAMINATION COULD NOT BE OBSERVED ON THE SHARED PICTURE AND AFTER THE INCUBATION, THE COMPLAINT CAN BE CONFIRMED BASED ON THE TREND THAT WAS IDENTIFIED. A THOROUGH INVESTIGATION IS CURRENTLY BEING PERFORMED. H3 OTHER TEXT : SEE H.10.
IT WAS REPORTED THAT WHILE USING PLATE COLUMBIA AG 5PRCT SB 90MM THE CUSTOMER OBSERVED BIOLOGICAL CONTAMINATION OF THE MEDIA. THIS EVENT OCCURRED 1 TIME(S). THERE IS NO INDICATION THAT FALSE RESULTS WERE REPORTED TO HEALTH CARE PRACTITIONERS. THERE IS NO INDICATION OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER FOUND A PLATE WITH CONTAMINATION, IDENTIFIED IN MALDITOF AS BACILLUS IDRIENSIS.
IT WAS REPORTED THAT WHILE USING PLATE COLUMBIA AG 5PRCT SB 90MM THE CUSTOMER OBSERVED BIOLOGICAL CONTAMINATION OF THE MEDIA. THIS EVENT OCCURRED 1 TIME(S). THERE IS NO INDICATION THAT FALSE RESULTS WERE REPORTED TO HEALTH CARE PRACTITIONERS. THERE IS NO INDICATION OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CUSTOMER FOUND A PLATE WITH CONTAMINATION, IDENTIFIED IN MALDITOF AS BACILLUS IDRIENSIS."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 742763 | PLATE COLUMBIA AG 5PRCT SB 90MM | CULTURE MEDIA, ENRICHED | KZI | BECTON DICKINSON GMBH | 1361180 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |