ID NOW COVID-19 ASSAY
Report
- Report Number
- 1221359-2022-02112
- Event Type
- Malfunction
- Date Received
- April 22, 2022
- Date of Event
- March 12, 2022
- Report Date
- November 28, 2022
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC
- Product Code
- QJR
- UDI-DI
- 10811877011320
- PMA / PMN Number
- EUA210264
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- 003
Narratives
THE INVESTIGATION IS STILL IN PROGRESS . A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. REFERENCE MFR. REPORTS: 1221359-2022-02111.
TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT M173879 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 191-000 / LOT M173879, TEST BASE PART NUMBER 190-430 / LOT M173879. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M173879 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE AS THE LOGFILES WERE NOT PROVIDED.
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE CORRECTIONS TO THE PREVIOUSLY REPORTED EVENT. PLEASE SEE UPDATES: D9, E3, G1, G4, G6 AND H2. CORRECTED DATA: G3- SUPPLEMENTAL REPORT 1 DATE RECEIVED BY MANUFACTURER :15-NOV-2021.
THE CUSTOMER REPORTED TWO FALSE POSITIVE RESULTS WITH ID NOW COVID-19 .THE SECOND TEST WAS PERFORMED ON (B)(6) 2022. CONFIRMATION TESTING WAS PERFORMED ON (B)(6) 2022, AND GENERATED NEGATIVE RESULT . NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED. NO INFORMATION WAS PROVIDED IF PATIENT WAS ASYMPTOMATIC OR SYMPTOMATIC . THIS MFR. REPORT ADDRESSES TEST TWO (2) OF TWO (2) TESTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 183818 | ID NOW COVID-19 ASSAY | MOLECULAR IVD FOR ID NOW COVID-19 | QJR | ABBOTT DIAGNOSTICS SCARBOROUGH, INC | M173879 | 10811877011320 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 3 YR | Female |