BD SYRINGE 50 ML CATHETER TIP
Report
- Report Number
- 1911916-2022-00212
- Event Type
- Malfunction
- Date Received
- April 22, 2022
- Date of Event
- March 30, 2022
- Report Date
- June 22, 2022
- Manufacturer
- BD MEDICAL (BD WEST) MEDICAL SURGICAL
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
THE FOLLOWING FIELDS HAVE BEEN UPDATED WITH ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL?: YES. RETURNED TO MANUFACTURER ON: 16-JUN-2022. H6: INVESTIGATION SUMMARY: IT WAS REPORTED THE PLUNGER ROD IS BROKEN. TO AID IN THE INVESTIGATION, ONE SAMPLE WITH NO PACKAGING BLISTER AND THREE PHOTOS WERE RECEIVED FOR EVALUATION BY OUR QUALITY TEAM. A VISUAL INSPECTION WAS PERFORMED AND THE PLUNGER ROD THUMB PRESS IS DAMAGED. THE THREE PHOTOS PROVIDED SHOW THE SAMPLE RECEIVED. THIS DEFECT COULD OCCUR IF THERE WAS A JAM DURING THE ASSEMBLY PROCESS INDUCING THE DAMAGE TO THE PLUNGER ROD. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 303553, LOT NUMBER 1103488. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. VERIFICATION OF THE ASSEMBLY PROCESS WAS PERFORMED. THE ALIGNMENT OF THE RAILS AND CONVEYORS WAS GOOD, AND THE FLOW OF PRODUCT WAS ACCEPTABLE. TO DATE, THERE HAVE BEEN NO OTHER SIMILAR EVENTS REPORTED FOR THIS LOT. BASED ON THE INVESTIGATION AND WITH THE RETURNED SAMPLE ANALYSIS THE SYMPTOM IS CONFIRMED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED BD SYRINGE 50 ML CATHETER TIP HAD A BROKEN PLUNGER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "SYRINGE'S PLUNGER IS BROKEN."
IT WAS REPORTED BD SYRINGE 50 ML CATHETER TIP HAD A BROKEN PLUNGER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "SYRINGE'S PLUNGER IS BROKEN."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 171225 | BD SYRINGE 50 ML CATHETER TIP | PISTON SYRINGE | FMF | BD MEDICAL (BD WEST) MEDICAL SURGICAL | 1103488 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |