FDA Adverse Event Malfunction Summary report: N

SALTER LABS

MDR report key: 14189567 · Received April 22, 2022

Report

Report Number
3000219639-2022-00022
Event Type
Malfunction
Date Received
April 22, 2022
Date of Event
March 14, 2022
Report Date
April 21, 2022
Manufacturer
SALTER LABS
Product Code
CAT
UDI-DI
00607411889097
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE IN QUESTION ALONE DOES NOT POSE A FIRE HAZARD BUT THE USE OF OXYGEN INCREASES THE RISK OF FIRE SPREADING THROUGH AN OXYGEN SUPPLY LINE. THE UNIT IN QUESTION IS EXPECTED TO MELT WHEN EXPOSED TO HEAT. THE IGNITION TEMPERATURE OF PVC IS ~851F THEREFORE MAKING IT UNLIKELY THAT THE TUBING COULD HAVE SPONTANEOUSLY COMBUSTED. OPEN FLAMES AND OR IGNITION SOURCES MAY INCLUDE CANDLES, CIGARETTES, CIGARS, FIRE PITS, BARBEQUES, ETC. OXYGEN SUPPLIERS ARE REQUIRED TO PROVIDE WARNINGS OF FIRE HAZARDS WHEN OXYGEN IS IN USE. (I.E., THE JOINT COMMISSION, NATIONAL PATIENT SAFETY GOALS, 2012). EVEN THOUGH THERE WAS NO PATIENT INJURY, OXYGEN TUBING CAUGHT ON FIRE DUE TO PATIENT WALKING WITH TUBING CLOSE TO THE CHIMNEY OUTLET. BASED ON THE INFORMATION ABOVE, THIS WOULD BE A REPORTABLE EVENT.

Additional Manufacturer Narrative · 0

THE DEVICE IN QUESTION ALONE DOES NOT POSE A FIRE HAZARD BUT THE USE OF OXYGEN INCREASES THE RISK OF FIRE SPREADING THROUGH AN OXYGEN SUPPLY LINE. THE UNIT IN QUESTION IS EXPECTED TO MELT WHEN EXPOSED TO HEAT. THE IGNITION TEMPERATURE OF PVC IS 851F THEREFORE MAKING IT UNLIKELY THAT THE TUBING COULD HAVE SPONTANEOUSLY COMBUSTED. OPEN FLAMES AND OR IGNITION SOURCES MAY INCLUDE CANDLES, CIGARETTES, CIGARS, FIRE PITS, BARBEQUES, ETC. OXYGEN SUPPLIERS ARE REQUIRED TO PROVIDE WARNINGS OF FIRE HAZARDS WHEN OXYGEN IS IN USE. (I.E., THE JOINT COMMISSION, NATIONAL PATIENT SAFETY GOALS, 2012). EVEN THOUGH THERE WAS NO PATIENT INJURY, OXYGEN TUBING CAUGHT ON FIRE DUE TO PATIENT WALKING WITH TUBING CLOSE TO THE CHIMNEY OUTLET. BASED ON THE INFORMATION ABOVE, THIS WOULD BE A REPORTABLE EVENT. A FIRE EVENT IS NOT RELATED TO A DEVICE MALFUNCTION. THE PRODUCTS ARE PERFORMING AS EXPECTED: A CONDUIT FOR OXYGEN DELIVERY. THE FIRE IS UNRELATED TO THE PRODUCT ITSELF - IT IS A BY-PRODUCT OF OXYGEN AND AN OPEN-SOURCE FLAME (E.G., SMOKING, CANDLES, SPARKS FROM OTHER SOURCES). IN THIS CASE, THERE WAS AN EMBER CLOSE TO THE TUBING AND WAS IGNITED DUE TO THE OXYGEN SOURCE. THE OXYGEN HAS SAFETY REQUIREMENTS DISPLAYED AND SAFETY SIGNS ARE PROVIDED. NO FURTHER INVESTIGATION IS REQUIRED. ONE OTHER COMPLAINT REGARDING FLAME HAS BEEN REPORTED IN THE LAST 12 MONTHS. PRODUCT WAS REPLACED. SAFETY INFORMATION REVIEWED WITH ACCOUNT BY OXYGEN SUPPLIER. THERE WAS NO MALFUNCTION OF THE SUNMED PRODUCT.

Description of Event or Problem · 0

INFLAMMATION OF THE TUBING.

Description of Event or Problem · 0

INFLAMMATION OF THE TUBING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1089083 SALTER LABS PATIENT KIT - (1) #1215, AND (1) #2025 (AIR LIQUIDE) - 25/CS CAT SALTER LABS SO-1188 UNKNOWN 00607411889097

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other