FDA Adverse Event Malfunction Summary report: N

TRUE METRIX

MDR report key: 14187732 · Received April 22, 2022

Report

Report Number
1000113657-2022-00231
Event Type
Malfunction
Date Received
April 22, 2022
Date of Event
March 28, 2022
Report Date
April 22, 2022
Manufacturer
TRIVIDIA HEALTH INC
Product Code
NBW
UDI-DI
0002129200783
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

INTERNAL REPORT REFERENCE NUMBER: (B)(4). METER AND TEST STRIPS WERE RETURNED FOR EVALUATION. PRODUCT TESTING WAS PERFORMED AND NO DEFECT FOUND ON RETURNED METER. TEST STRIPS ARE EXPIRED - UNABLE TO TEST. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-012: PRODUCT EXPIRED. NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL ON 11-APR-2022 TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - ABLE TO ESTABLISH CONTACT WITH CUSTOMER WHO STATED REPLACEMENT PRODUCTS RESOLVED INITIAL CONCERN.

Description of Event or Problem · 0

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 201, 268, 203, 269 AND 341 MG/DL. THE CUSTOMER¿S EXPECTED AM FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 100-130 MG/DL AND EXPECTED PM NON-FASTING BLOOD GLUCOSE TEST RESULT IS 150 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE TEST STRIPS ARE EXPIRED: MANUFACTURER¿S EXPIRATION DATE IS 07/31/2021; OPEN VIAL DATE AND STORAGE WERE NOT DISCLOSED. THE CUSTOMER DID NOT HAVE ANOTHER VIAL OF TEST STRIPS THAT HAD BEEN STORED AND HANDLED CORRECTLY. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
328941 TRUE METRIX SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER NBW TRIVIDIA HEALTH INC STRIP, TMX CVS 100CT12/CASE MG/DL MW4099S 0002129200783

Patients

Seq Age Sex Outcome Treatment
1 Unknown