DSL-10-2000 ACTIVE® CORTISOL ENZYME IMMUNOASSAY (EIA) KIT
Report
- Report Number
- 1628193-2009-00001
- Event Type
- Malfunction
- Date Received
- July 21, 2009
- Date of Event
- July 9, 2009
- Report Date
- July 21, 2009
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- CGR
- PMA / PMN Number
- K884689
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
ROOT CAUSE OF THE SHIFT IN THE CALIBRATION OF THE ASSIGNED STANDARD VALUES IS CURRENTLY UNDER INVESTIGATION. A MALFUNCTION WILL BE ASSUMED FOR THE PURPOSE OF THIS REPORT.
DURING AN INTERNAL TESTING CONDUCTED BY BECKMAN COULTER INC. (BCI), IT WAS REPORTED THAT FALSE HIGH CORTISOL VALUES WERE GENERATED BY THE DSL-10-2000 ACTIVE CORTISOL ENZYME IMMUNOASSAY (EIA) KIT. BECKMAN COULTER HAS CONFIRMED THAT THE FOLLOWING LOT NUMBERS: 04195, 04195A, 04195B, 04195D, 04195E, 04195F, 09266A, 09266, 04126A, 04126, 03066A, 03066, 01036, 716217, 716218, 800408, 800409, 800410, 890929 AND 990168 OF ACTIVE CORTISOL ELISA CALIBRATORS OVER ESTIMATE THE EXPECTED DOSE VALUES FOR PATIENT SERUM SAMPLES DUE TO A SHIFT IN THE ASSIGNED CALIBRATOR VALUES. AN EXAMPLE FOR LOT 990168 WHICH YIELDED VALUES SIGNIFICANTLY ABOVE (2-5X) OTHER PRODUCTS. ALL KITS WITH AFFECTED LOTS RESULTED IN OVERESTIMATION OF CORTISOL DOSE. PATIENT RESULTS WERE NOT GENERATED IN THIS EVENT SINCE THE EVENT WAS DISCOVERED IN-HOUSE AND NO PATIENT SAMPLES WERE TESTED AT THE TIME. THERE WAS NO AFFECT TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DSL-10-2000 ACTIVE® CORTISOL ENZYME IMMUNOASSAY (EIA) KIT | ACTIVE® CORTISOL EIA, 96 WELLS | CGR | BECKMAN COULTER INC. | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |