FDA Adverse Event Other Summary report: N

DA VINCI S SURGICAL SYSTEM

MDR report key: 1416635 · Received April 24, 2009

Report

Report Number
2955842-2009-00143
Event Type
Other
Date Received
April 24, 2009
Date of Event
April 1, 2009
Report Date
April 1, 2009
Manufacturer
INTUITIVE SURGICAL, INC.
Product Code
NAY
PMA / PMN Number
K081207
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

RESULTS AND CONCLUSIONS - THE INVESTIGATION CONDUCTED BY AN ISI FIELD SERVICE ENGINEER (FSE) CONFIRMED THE CUSTOMER REPORTED FAILURE AS THE ASSISTANT MONITOR DID NOT DISPLAY ANY VIDEO AND THE MESSAGE "COMPOSITE SEARCHING" WAS DISPLAYED. THE FSE SWITCHED THE MONITOR SETTING TO SDI VIDEO; HOWEVER, THERE WAS NO IMAGE DISPLAY. REVIEW SYSTEM ERROR LOGS FOR THE DAY DID NOT REVEAL ANY UNUSUAL ERRORS. EXAMINATION OF THE CONNECTIONS AND SETTINGS ON THE VSC LIGHT FOUND THAT THE GREEN LIGHT FOR THE DIGITIZER WAS NOT PRESENT. REMOVAL OF THE SIDE CONNECTIONS REVEALED THAT THE SITE'S ELECTROSURGICAL UNIT (ESU) WAS PRESSED AGAINST 3 PLUGS ON THE ISOLATION TRANSFORMER AND WERE UNSEATED. ONE OF THE AFFECTED PLUGS WAS THE DIGITIZER POWER CORD. THE FSE CONCLUDED THAT ESU PRESSED AGAINST THE 3 PLUGS CAUSING THE LOSS OF POWER. THE VIDEO WAS RESTORED TO THE ASSISTANT AND EXTERNAL MONITORS WHEN THE ESU AND PLUGS WERE RESEATED. AS OF (B)(6) 2009 THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL.

Description of Event or Problem · 1

IT WAS REPORTED THAT APPROXIMATELY 20 MINUTES INTO A DA VINCI S GYNECOLOGICAL SURGICAL PROCEDURE WHILE DOCKING THE SYSTEM, THE VISION SIDE CART (VSC) TOUCH SCREEN IMAGE WENT BLACK WITH THE MESSAGE "SDI SEARCHING" DISPLAYED IN THE LOWER RIGHT CORNER. THE EXTERNAL MONITOR CONNECTED TO THE CONSOLE ALSO WENT BLACK. WITH ASSISTANCE OF A TECHNICAL SUPPORT ENGINEER (TSE) THE SITE MOVED THE TOUCH SCREEN THROUGH IT'S RANGE OF MOTION, CHECKED THE VIDEO CABLES LOCATED AT THE REAR OF THE DIGITIZER AND THE CONSOLE, CONNECTED THE COMPOSITE CABLE TO THE DIGITIZER, CONNECTED ANOTHER MONITOR TO THE COMPOSITE CABLE, AND TRIED THE VSC AND CONSOLE CONNECTION; HOWEVER, THE IMAGE DID NOT RETURN. THE SURGEON DECIDED TO CONVERT TO TRADITIONAL OPEN SURGICAL TECHNIQUES TO COMPLETE THE PLANNED PROCEDURE. NO PT HARM WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DA VINCI S SURGICAL SYSTEM ENDOSCOPIC INSTRUMENT CONTROL SYSTEM NAY INTUITIVE SURGICAL, INC. IS2000 A5.1P7

Patients

Seq Age Sex Outcome Treatment
1 Other DA VINCI S SURGICAL SYSTEM| ACCESSORIES| INSTRUMENTS