FDA Adverse Event Injury Summary report: N

BALL HEADS: COCR COCR BALL HEAD 12/14 Ø 22 SIZE M 0

MDR report key: 14165804 · Received April 21, 2022

Report

Report Number
3005180920-2022-00274
Event Type
Injury
Date Received
April 21, 2022
Date of Event
March 19, 2022
Report Date
April 21, 2022
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JDI
UDI-DI
07630030805479
PMA / PMN Number
K080885
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 20 APRIL 2022. LOT 2010474: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 26-JAN-2021. EXPIRATION DATE: 2026-01-13. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT. ADDITIONAL IMPLANT INVOLVED: BALL HEADS: BIPOLAR HEAD 25060.2241 BIPOLAR HEAD Ø22X41 (K091967) LOT. 1905687. BATCH REVIEW PERFORMED ON 20 APRIL 2022. LOT 1905687: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 23-OCT-2019. EXPIRATION DATE: 2024-10-12. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT.

Description of Event or Problem · 0

AT ABOUT 1 MONTH AFTER THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE COCR HEAD AND BIPOLAR HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1958646 BALL HEADS: COCR COCR BALL HEAD 12/14 Ø 22 SIZE M 0 METALLIC BALL HEAD JDI MEDACTA INTERNATIONAL SA 01.25.123 2010474 07630030805479

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention