FDA Adverse Event Malfunction Summary report: N

HD AUTOCLAVABLE CAMERA HEAD

MDR report key: 14161264 · Received April 20, 2022

Report

Report Number
8010047-2022-06694
Event Type
Malfunction
Date Received
April 20, 2022
Report Date
June 9, 2022
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
FET
UDI-DI
04953170310799
PMA / PMN Number
K102059
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS EVALUATED BY OLYMPUS. THE EVALUATION FOUND RUST ON THE SHAFT OF THE COUPLER. ADDITIONAL EVALUATION FINDINGS ARE AS FOLLOWS: CONTROL UNIT APPEARANCE DEFECTS; AIR/WATER LEAKAGE; IMAGE QUALITY ABNORMAL DUE TO WHITE DOT. THE INVESTIGATION IS ONGOING AND THE DEFINITIVE ROOT CAUSE OF THE REPORTED EVENT CANNOT BE DETERMINED AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE SUPPLEMENTED ACCORDINGLY.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE APPROVED FINAL INVESTIGATION. CORRECTION TO G3 OF THE INITIAL MEDWATCH. THE AWARE DATE SHOULD BE 22-MARCH-2022. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. BASED ON THE RESULTS OF THE INVESTIGATION, THE DEFINITIVE ROOT CAUSE OF THE CORROSION COULD NOT BE DETERMINED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

A USER FACILITY RETURNED THE OLYMPUS, MODEL CH-S190-XZ-E, HD AUTOCLAVABLE CAMERA HEAD TO OLYMPUS FOR REPAIR DUE TO A REPORT OF A "CRACK ON CAMERA HEAD." UPON INSPECTION AND TESTING OF THE RETURNED DEVICE, RUST WAS FOUND ON THE SHAFT OF THE COUPLER. THIS REPORT IS SUBMITTED DUE TO THE RUST FOUND ON THE SHAFT OF THE COUPLER. AS THE PROBLEM WAS FOUND DURING IN-HOUSE SERVICE OF THE DEVICE, THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2719576 HD AUTOCLAVABLE CAMERA HEAD AUTOCLAVABLE CAMERA HEAD FET OLYMPUS MEDICAL SYSTEMS CORP. CH-S190-XZ-E 04953170310799

Patients

Seq Age Sex Outcome Treatment
1 Unknown