FDA Adverse Event Malfunction Summary report: N

FLEXIBLE HINGE TOE IMPLANT W/ GROMMETS #2S

MDR report key: 14160654 · Received April 20, 2022

Report

Report Number
3010667733-2022-00164
Event Type
Malfunction
Date Received
April 20, 2022
Date of Event
March 22, 2022
Report Date
June 13, 2022
Manufacturer
WRIGHT MEDICAL TECHNOLOGY INC
Product Code
KWH
UDI-DI
00840420127471
PMA / PMN Number
K780132
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

DEVICE WILL NOT BE RETURNED. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, IT WILL BE PROVIDED ON A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 0

CORRECTION D9/H3, H6 METHOD CODE; THE REPORTED EVENT COULD BE CONFIRMED, SINCE THE DEVICE WAS RETURNED WITH THE PROXIMAL GROMMET MISSING AND TWO DISTAL GROMMETS FOUND INSTEAD IN THE PACKAGING. BASED ON INVESTIGATION, THE ROOT CAUSE WAS ATTRIBUTED TO A MANUFACTURING RELATED ISSUE. THE FAILURE WAS CAUSED BY IMPROPER HANDLING/PACKAGING OF DEVICE. THE CURRENT INSTRUCTIONS FOR USE STATES: "THE SWANSON FLEXIBLE HINGE TOE JOINT GROMMET IS A THIN TITANIUM SHIELD DESIGNED FOR USE WITH THE SWANSON FLEXIBLE HINGE TOE IMPLANT IN RHEUMATOID PATIENTS WHERE CUTTING OR ABRASION OF THE FLEXIBLE IMPLANT FROM CONTACT WITH THIN, SHARP BONE EDGES CAN OCCUR, OR IN PATIENTS WHO HAVE HIGH ACTIVITY LEVELS. IT IS CONTOURED TO CONFORM TO THE SHAPE OF THE MIDSECTION OF THE FLEXIBLE IMPLANT AND IS FABRICATED FROM UNALLOYED TITANIUM FOR SURGICAL APPLICATION. THE DISTAL GROMMET IS USED ON THE DISTAL STEM AND THE PROXIMAL GROMMET IS USED ON THE PROXIMAL STEM TO PROTECT THE IMPLANT FROM BIOMECHANICAL SHEARING FORCES OF SHARP BONE EDGES DURING JOINT MOTION." A NONCONFORMANCE WAS OPENED INTO THIS ISSUE TO FURTHER INVESTIGATE THE PROBLEM.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE SWANSON IMPLANT BOX WAS OPENED, TWO PROXIMAL GROMMETS IN PACKAGE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE SWANSON IMPLANT BOX WAS OPENED, TWO PROXIMAL GROMMETS IN PACKAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1186094 FLEXIBLE HINGE TOE IMPLANT W/ GROMMETS #2S PROSTHESIS, TOE, CONSTRAINED, POLYMER KWH WRIGHT MEDICAL TECHNOLOGY INC G4260102 1717547 00840420127471

Patients

Seq Age Sex Outcome Treatment
1 Unknown