LEEP PRECISION
Report
- Report Number
- 1216677-2022-00114
- Event Type
- Malfunction
- Date Received
- April 20, 2022
- Date of Event
- April 4, 2022
- Report Date
- May 27, 2022
- Manufacturer
- COOPERSURGICAL, INC.
- Product Code
- HGI
- PMA / PMN Number
- K963348
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
COOPERSURGICAL, INC. IS INVESTIGATING THE REPORTED CONDITION.
INVESTIGATION. X-REVIEW DHR. X-INSPECT RETURNED SAMPLES. *ANALYSIS AND FINDINGS COMPLAINT (B)(4). DISTRIBUTION HISTORY: THIS COMPLAINT UNIT WAS MANUFACTURED AT CSI ON 01/28/2021 UNDER WO#'S (B)(4) & (B)(4) AND SHIPPED ON 02/21/2021. MANUFACTURING RECORD REVIEW: DHR'S (B)(4) & (B)(4) WERE REVIEWED AND NO NON-CONFORMITIES, RELATED TO THE COMPLAINT CONDITION, WERE NOTED. INCOMING INSPECTION REVIEW: NOT APPLICABLE. SERVICE HISTORY RECORD: THIS UNIT WAS REPAIRED UNDER A RE-WORK WORK ORDER (B)(4) JANUARY 2022 AND RETURNED FEBRUARY 2022. THE REPAIR IS UNRELATED TO THE COMPLAINT. HISTORICAL COMPLAINT REVIEW: A REVIEW OF THE 2-YEAR COMPLAINT HISTORY SHOWED SIMILAR REPORTED COMPLAINT CONDITION. PRODUCT RECEIPT: THE COMPLAINT UNIT WAS RETURNED ON A REPAIR. HOWEVER, BASED ON LOG (B)(4), THIS UNIT WAS AT CSI ON 4/12/2022. VISUAL EVALUATION: VISUAL EXAMINATION OF THE COMPLAINT UNIT REVEALED NO PHYSICAL DAMAGE. FUNCTIONAL EVALUATION: COMPLAINT UNIT WAS FUNCTIONALLY EVALUATED AND FOUND NOT TO FUNCTION PROPERLY. THE UNIT FAILED THE RF LEAKAGE TEST SECTION FOR COAG OF THE STANDARD IN-PROCESS TESTING. *WAS THE COMPLAINT CONFIRMED? YES. ROOT CAUSE: A ROOT CAUSE FOR THE COMPLAINT CONDITION IS NOT DEFINITIVELY DETERMINED. THE COAG FUNCTION IS KNOWN REQUIRE A SITE THAT HAS MINIMAL BLOOD PRESENT AND MAY NOT WORK WELL IN AN EXCESSIVELY WET ENVIRONMENT. IT IS NOT CLEAR IF THE LEAKAGE ERROR DETECTED IS A SOLE CONTRIBUTING FACTOR OR IF THE EXCESSIVE BLEEDING NOTED WAS A EQUAL OR GREATER CONTRIBUTOR TO THE COMPLAINT CONDITION. BASED ON THE INFORMATION AVAILABLE THIS FINDING SUGGESTS A POTENTIAL FOR A DESIGN RELATED ISSUE BEST EVALUATED BY TECH OPS. CORRECTION AND/OR CORRECTIVE ACTION . THE UNIT WAS FITTED WITH A NEW BOARD, TESTED TO SPECIFICATIONS AND RETURNED TO THE CUSTOMER. THE MAIN BOARD PULLED OUT (S/N (B)(6)) WAS SET ASIDE FOR FURTHER EVALUATION BY THE TECH OPS EE UNDER PROJECT #(B)(4). COOPERSURGICAL WILL CONTINUE TO MONITOR THIS COMPLAINT CONDITION FOR TRENDS.
2ND INCIDENT IN REFERENCE TO (B)(4). COMPLAINT INFORMATION PROVIDED BY CSI SALES: ON (B)(6) 2022, DR. (B)(6) 26 YO FEMALE HGSIL (CIN III) GYNEX RADIUS LOOP ELECTRODE GYNEX BALL ELECTRODE MEDLINE EXCESSIVE BLEEDING/CAUTERY MISFIRE. PROVIDER USED THE CUT SETTING TO MAKE THE INITIAL INCISION. PROVIDER USE A 50:50 MIX OF LIDOCAINE 1% AND LIDOCAINE WITH EPI FOR CERVICAL BLOCKS TO HELP MANAGE BLEEDING. PATIENT REQUIRED MEDICAL ATTENTION WITH PACKING AND LYING DOWN UNTIL BLEEDING COULD BE CONTROLLED. ON (B)(6) 2022- UPDATED WITH REPAIR LOG # 98250. PER SERVICE & REPAIR TEAM- RECEIVED LEEP PRECISION HERE ON DUAL COMPLAINT FOR EXCESSIVE BLEEDING. IT IS ON REPAIR LOG 98250. EMERGENCY DEPARTMENT DID A FULL TEST AND FINDS THE RF LEAKAGE FOR COAG IS OUT OF SPEC. NO OTHER ISSUES. LEEP PRECISION INTG SYS LP-10-120 (B)(4).
2ND INCIDENT IN REFERENCE TO E-COMPLAINT (B)(4). COMPLAINT INFORMATION PROVIDED BY CSI SALES- (B)(6) 2022-DR. (B)(6) ((B)(6) -26 YO FEMALE HGSIL (CIN III) GYNEX RADIUS LOOP ELECTRODE GYNEX BALL ELECTRODE MEDLINE EXCESSIVE BLEEDING/CAUTERY MISFIRE. PROVIDER USED THE CUT SETTING TO MAKE THE INITIAL INCISION. PROVIDER USE A 50:50 MIX OF LIDOCAINE 1% AND LIDOCAINE WITH EPI FOR CERVICAL BLOCKS TO HELP MANAGE BLEEDING. PATIENT REQUIRED MEDICAL ATTENTION WITH PACKING AND LYING DOWN UNTIL BLEEDING COULD BE CONTROLLED. (B)(6) 2022- UPDATED WITH REPAIR LOG # (B)(4). PER SERVICE & REPAIR TEAM- RECEIVED LEEP PRECISION HERE ON DUAL COMPLAINT FOR EXCESSIVE BLEEDING. IT IS ON REPAIR LOG (B)(4). ED DID A FULL TEST AND FINDS THE RF LEAKAGE FOR COAG IS OUT OF SPEC. NO OTHER ISSUES. 1216677-2022-00114 LEEP PRECISION INTG SYS LP-10-120 E-COMPLAINT-(B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 52444 | LEEP PRECISION | LEEP PRECISION | HGI | COOPERSURGICAL, INC. | LP-10-120 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |