FDA Adverse Event Malfunction Summary report: N

ENDOPATH THORACIC ENDO LINEAR CUTTER W/SAFETY LOCK

MDR report key: 141441 · Received December 31, 1997

Report

Report Number
1527736-1997-03682
Event Type
Malfunction
Date Received
December 31, 1997
Date of Event
December 4, 1997
Report Date
December 4, 1997
Manufacturer
ETHICON ENDO-SURGERY, INC.
Product Code
GAG
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT COMPLAINT ANALYSIS TEAM HAS COMPLETED ITS INVESTIGATION OF THE DEVICE WHICH WAS RETURNED. THE RESULTS OF THE INVESTIGATION CONDUCTED BY THE APPROPRIATE ENGINEERS AND TECHNICIANS ARE LISTED BELOW. VISUAL INSPECTIONS & RESULTS: LOCKOUT POSITION, NA; CARTRIDGE CONDITION, NA AND CARTRIDGE RETURN BATCH NUMBER, NA. FUNCTIONAL TESTS & RESULTS: CONDITION OF FIRING TRIGGER LOCKOUT, GOOD; CONDITION OF PINION GEAR, GOOD; CONDITION OF SHORT RACK, GOOD; CONDITION OF YOKE, GOOD AND TEST CARTRIDGE BATCH #, K01142. ANALYSIS CONCLUSION: BASED UPON THE INQUIRY INFO REC'D, THE VISUAL EXAMINATION, AND THE FUNCTIONAL TESTING, NO CONCLUSION COULD BE REACHED AS TO WHY THE INSTRUMENT REPORTEDLY "DID NOT CUT" DURING SURGERY. THE INSTRUMENT WAS RETURNED IN GOOD PHYSICAL CONDITION. THE INSTRUMENT WAS CYCLED, FIRED, CUT, AND FORMED THE STAPLES WITHIN DESIGN SPEC. THE INSTRUMENT WAS DISASSEMBLED TO EXAMINE THE INTERNAL COMPONENTS AND NO DEFORMATIONS COULD BE IDENTIFIED. IT WAS CONCLUDED THAT THE INSTRUMENT WAS FULLY FUNCTIONAL AND CONFORMING TO DESIGN SPECS. THE EXPERIENCE THE SURGEON REPORTED COULD NOT BE REPEATED. EACH INSTRUMENT IS EVALUATED DURING THE ASSEMBLY PROCESS TO ENSURE IT FUNCTIONS PROPERLY.

Description of Event or Problem · 1

IT WAS REPORTED DURING A THORACOSCOPY AFTER THE THIRD FIRING OF THE EZ45B THE INSTRUMENT PLACED STAPLES BUT DID NOT APPEAR TO CUT THE LUNG TISSUE. ENDOSCISSORS WERE USED TO TRANSECT BETWEEN STAPLE LINES. PROCEDURE CONTINUED WITH A NEW EZ45B. THERE WAS NO PT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH THORACIC ENDO LINEAR CUTTER W/SAFETY LOCK ENDOSCOPIC LINEAR CUTTER GAG ETHICON ENDO-SURGERY, INC. NA K4668V

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other