CHECK-IT
Report
- Report Number
- 3016521623-2022-00160
- Event Type
- Malfunction
- Date Received
- April 19, 2022
- Date of Event
- April 12, 2022
- Report Date
- April 27, 2022
- Manufacturer
- LUCIRA HEALTH, INC.
- Product Code
- QJR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
PRODUCT HAS NOT BEEN RETURNED. A DHR REVIEW OF THE ASSOCIATED KIT LOT NUMBER (K08A112302224M1) WAS COMPLETED, AND 0 DISCREPANCIES WERE FOUND (ALL LOTS PASSED IN TOTAL PER THE SAMPLING PLAN). A REVIEW OF EXISTING CAPAS, SCARS AND NCMRS WERE COMPLETED AND THERE ARE NO PRIOR RECORDS RELATED TO "FALSE POSITIVE" FAILURE MODE FOR THIS LOT NOR ANY PREVIOUS COMPLAINTS FROM THIS CUSTOMER ASSOCIATED WITH "FALSE POSITIVE" RESULTS PRIOR TO RECEIVED DATE OF (B)(6) 2022. BASED ON REVIEW OF THE PRODUCT'S FMEAS AND RISK ASSESSMENT DOCUMENTATION, FALSE POSITIVE TEST RESULTS ARE A KNOWN POSSIBLE OUTCOME REGARDING THIS ISSUE UNDER EVALUATION, REFER TO FMEA-001 AND FMEA-004. (B)(4). LUCIRA HEALTH WILL CONTINUE TO MONITOR TRENDS RELATED TO FALSE POSITIVE RESULTS IN ACCORDANCE WITH POST-MARKET SURVEILLANCE PROCESS. DHR REVIEW KIT LOT# K08A112302224M1: SAMPLE VIAL LOT DHR REVIEW: 2202226, 2202228 (ASSOCIATED INTERNAL LOT #220124-1B, 220124-1D) TEST LOT DHRS REVIEW: 2202235, 2202061, 2202266, 2202234, 2202235, 2202268, 2202236 (ASSOCIATED INTERNAL LOT # 220124-12R, 220008-20O, 220008-20U, 220124-12P, 220124-12T, 220008-23B, 220124-12W) . BASED ON THE LIMITED INFORMATION AVAILABLE, ROOT CAUSE CANNOT BE DETERMINED. POTENTIAL ROOT CAUSES INCLUDE BUT ARE NOT LIMITED TO: ENVIRONMENTAL CONTAMINATION, LOW VIRAL LOAD, AND/OR DEVICE FAILURE. A SUPPLEMENTAL REPORT WILL BE FILED IF ANY FURTHER INVESTIGATION AND/OR ADDITIONAL INFORMATION IS OBTAINED. BASED ON THE COMPLAINT REVIEW, NO HARM WAS REPORTED WITHIN THE COMPLAINT. THIS DEVICE IS MARKETED UNDER EUA (B)(4) CHECK-IT.
PRODUCT HAS NOT BEEN RETURNED. A DHR REVIEW OF THE ASSOCIATED KIT LOT NUMBER: (K08A112302224M1) WAS COMPLETED, AND 0 DISCREPANCIES WERE FOUND (ALL LOTS PASSED IN TOTAL PER THE SAMPLING PLAN). A REVIEW OF EXISTING CAPAS, SCARS AND NCMRS WERE COMPLETED AND THERE ARE NO PRIOR RECORDS RELATED TO "FALSE POSITIVE" FAILURE MODE FOR THIS LOT NOR ANY PREVIOUS COMPLAINTS FROM THIS CUSTOMER ASSOCIATED WITH "FALSE POSITIVE" RESULTS PRIOR TO RECEIVED DATE OF APRIL 12, 2022. (B)(4). DHR REVIEW KIT LOT#: K08A112302224M1: SAMPLE VIAL LOT DHR REVIEW: 2202226, 2202228 (ASSOCIATED INTERNAL LOT#: 220124-1B, 220124-1D). TEST LOT DHRS REVIEW: 2202235, 2202061, 2202266, 2202234, 2202235, 2202268, 2202236 (ASSOCIATED INTERNAL LOT # 220124-12R, 220008-200, 220008-20U, 220124-12P, 220124-12T, 220008-23B, 220124-12W). A SUPPLEMENTAL REPORT WILL BE FILED IF ANY FURTHER INVESTIGATION AND/OR ADDITIONAL INFORMATION IS OBTAINED. BASED ON THE COMPLAINT REVIEW, NO HARM WAS REPORTED WITHIN THE COMPLAINT. THIS DEVICE IS MARKETED UNDER EUA: (B)(4) CHECK-IT.
ONE DEVICE REPORTED AS HAVING A FALSE POSITIVE RESULT. DURING A (B)(6) SERVICES KICKOFF MEETING (LUCIRA HEALTH HELPED TO FACILITATE), THERE WERE 157 LUCIRA TEST KITS SELF-ADMINISTERED. AS A RESULT, AN ALLEGED FALSE POSITIVE RESULT WAS IDENTIFIED FROM 1 LUCIRA TEST KIT (ONE INDIVIDUAL). THE INDIVIDUAL TOOK 2 ANTIGEN TESTS WITH NEGATIVE RESULTS FOR CONFIRMATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 986937 | CHECK-IT | COVID-19 TEST KIT | QJR | LUCIRA HEALTH, INC. | K08A112302224M1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |